Advanced Medical Equipment

    Equipment & Furniture Component  
  LIST OF MEDICAL EQUIPMENT  
       
Item Name Quantity Technical Specifications and Standards
No        
       
      Processing machine for X-ray films from 13×18 cm. 18×24 cm.24×30
      cm.30x40cm.35.6×35.6 cm. 35x43cm.  
      Capacity at least 60 films per hour  
  X-Ray Film processor   Standard Equipment: Processor + set of 3 replenish bottles  
1   Accessories: Water supply connection with shut-off valve , Filter , Light-tight
Table top  
  1 cover , Processor stand  
     
      Automatic stand-by mode.  
      Automatic film detection  
      Power Requirements :  
      220 VAC ±10 % 50 Hz .  
         
      X-ray Film Illuminator (Viewer) . 2 Fields . Wall-Mount  
      Diffuse, uniform, flicker free illumination  
      Transparent Spring-loaded film Retainers shall grip lightly and firmly without
      obscuring top edge details  
      Screen shall be recessed into the cabinet to help keep the interior dust free and
      eliminate side light spill  
      The film retainers shall always operate effectively maintaining an even pressure
2 X-ray Film Viewer   across the full width of the illuminator.  
  Direct Starting , Luminous Source : Daylight  
       
      Command and Control on Front Side.  
      Bipolar Switch with Pilot lamp  
      Wall Fixations shall be provided for . Cord , Local Plug  
    8 Power Supply : 220V/50Hz  
      CE;  
      EC Marked  
      US FDA;  
      ISO certification  
      Universal, Remote controlled Universal Unit  
      Processor controlled, Mixed Cassette’s Radiography and Digital Fluoroscopy
      screening System.  
      Overhead Tube on the freely moving Tube Arm without floor mounted column.
      The System to be suitable for Standard Skeletal and Radiographic Examinations,
      Including Lateral Exposures and Oblique Beam Projections
      Fully Automatic Under Table Spot Film Device for Cassette Radiography with
    1 Extensive Range of Cassettes, Grids and Attachments for Image Intensifier.
      Free Cassette Exposures on Table, Floor, Wheelchair or Gurney.
      Table.  
      Tilt: Motor Driven, +90º to -15º, automatic stop in horizontal position
      Height: approximately 85 cm.  
      Tabletop Outside Dimensions: length approximately 200cm,Width approximately
      80cm  
3 X-Ray Unit Universal   Radiolucent: not less than 190cm x 55cm.  
      Longitudinal Travel: Motor Driven, at least 160 cm  
      Transverse Travel: at least 20 cm  
      Patient Weight: at least 200 kg.  
      Tube Assembly.  
      Max. exposure voltage : 150 kV  
      Anode heat dissipation rate at least 30 kW for small focus and 60 kW for large
      focus  
      Anode heat storage capacity at least 500 kHU  
      Dual focus 0.6 mm and 1 mm  
      Complete filtration W ≥ 2.5 mm Al  
      Focus-Film Distance (SID) Fixed Variable : 115cm / 150cm. 115cm and 150cm
      must be set by Motor Driven with Adjustment Speed  
      Oblique Projections: max.-40º-+40º ( SID 115cm.) and -35º-+35º (SID 150cm),
      Automatic Parallax Compensation Between Cassette and Image Intensifier Input
      Screen in Central Ray.  

 

page 1 of 27

 

Equipment & Furniture Component

Tube Assembly Swivel: Manually in the Range +90º to -90º with Stops max. every 10º and -90º to 180º with stops max. every 30º.

Spot Film Device.

Front loading, automatically drawing in and out, centering and format sensing for

Cassettes, of the Formats 18x24cm (8″x10″) to 35 x 43cm (14″x17″), Standardized

 

According to IEC, ANSI and DIN

Automatic Format Collimation: must be Selected Separately According to Format

Height/Width for Cassette Spot filming; Automatic Formatting for Bucky

Exposures, Object-related Collimation.

 

Film Segmentation: Minimum 4 on 1., in Multiple Segmentation According to

Cassettes Program

Inward Movement Time: Park to Exposure position not more than 1sec.

Time Interval: Fluoroscopy/Radiography ≤1.0 sec. without grid movement.

 

Scattered Radiation Grid: Stationary 17:1, 70 Lines/cm, Fo=125cm.,

Excursion: minimum 105cm, Remote-controlled.

Compression Device.

 

Radio transparent, Remotely Controlled, Detachable and Replaceable Cone minimum 3 Shapes.

 

Compression Force range: from 5 to 155N, Movement Blockage starting from 50N Force Indication: Digital at the System Remote Control Console.

Projection Angle: -30º-+30º.

 

Image Intensifier and TV System

3 Image Fields 13 to 23 cm

Visual resolution: Mean Value 5.2; 5.8; 6.4 Lp/mm

 

Television System Standard: Line Frequency 50Hz; 625 Lines with 50Hz. TV Technology : CCD sensor, noise suppression, LIH function Automatic Dose Stabilization (ADR)

 

Monitor: 44cm or more, Frame Rate at least 50 Hz.

 

X-Ray Generator

 

High Frequency, Multipulse Up to 150 kV, ≥50kW.

 

Automatic System: 1-Point Technique with Continuously Falling Load; 2 -,3- Point Technique with Constant Load; 3-Point Technique with AEC

 

Organ Programs: minimum 20 Organ Programs must be generated and stored. Number of Workstations: not less than 5 of with 1 Fluoroscopic Workstation Fluoroscopy working range: minimum 40kV/ 0.5mA up to maximum 110kV≤10mA.

 

Fluoroscopic Mode Selectable: Fluoroscopy Under Manual Control / Automatic Control

 

Fluoroscopy Under Manual Control: manual select kV but mA must be calculated from Anti-isowatt, curve

 

Fluoroscopy Under Automatic Control: Minimum 2 required Characteristic Curve must be selected for Automatic Dose Rate Control : us Minimal Dose/ Contrast Litho ( or other according customer ordering)

 

Tube Connection: minimum 2 Double-Focus Tube Assembly. 3 field Automatic Exposure Control

 

Primary Collimator: Inherent Filtration 1.0mm Al; additional Filters 0.1mm Cu, 0.2mm Cu, 0.3mm Cu.

Ambient Conditions

 

Operation: +10Cº to 35Cº; 20% to 75% relative humidity, non-condensing; 70kPa to 106kPa

 

Storage and Transport: -20Cº to+70Cº; 10% to 95% relative humidity, non-condensing; 70kPa to 106kPa

Power Connection

 

Nominal Voltage: 3/N/PE, 380V±10%

Nominal Frequency: 50Hz.

Completeness:

 

Lead glass, radiation protection window 80x100cm. ±10 % , 2,1 Pb equiv X-ray protection:

 

1x Coat apron, 0.35 Pb, size medium, 110cm. X-ray protection coat apron for general examinations.

 

1x Gonad protection apron 40x37cm, 0.5 Pb. Apron for adults for protection of ovarian and gonad.

 

1x Gonad protection capsule for boys, 1.0 Pb.

  • Set of cassettes
  • Cassette without patient ID window
  • Fitted with green screen

2x Cassette with green screen, 18 cm x 24 cm

 

 

page 2 of 27

 

        Equipment & Furniture Component    
                 
            1x Cassette with green screen, 18 cam 43 cm    
            2x Cassette with green screen, 24 cm x 30 cm    
            2x Cassette with green screen, 35 cm x 43 cm    
            1x Cassette with green screen, 35 cm x 35 cm    
            Exposure range at least 0.5–200 mAs    
            Shortest exposure time: not more than 10ms.    
            Max. exposure time: not less than 5 sec.    
            Tube head must be rotatable in all planes.    
            Adjustable collimator and light marker Easy to use and to move. Non automotive  
            Two point technique – kV and mAs.    
            High- frequency X-ray generator with multipulse voltage provider.  
4   X-Ray Mobile   1   The touch-screen keys and digital display of kV and mAs values.  
            Output: not less than 2.5kW. 120 kV, 40mA.    
            Focal spot not more than 1 mm.    
            Collimator: rotatable ± 90°    
            Max. horizontal extension at least 800mm    
            Power Requirements    
            220 V ±10%, 50Hz    
            Wide Dynamic Range Digital Front-end    
            The 12-bit A/D converter    
            Multi-beam processing.    
            Image Display Modes    
            B: gray-scale imaging, M,D: Spectral Doppler (PW,HPRF PW),  
            Flow: Color Doppler and Power Flow imaging, Dual B, B and M, B and D, B(Flow),  
            Dual B (Flow), M(Flow), B(Flow) and M(Flow), B(Flow) and D,  
            Triplex mode: B, Flow and PW Doppler simultaneous real-time display,  
            Dual Dynamic Display (DDD): B and B (Flow) simultaneous real-time display, TDI  
            (Tissue Doppler Imaging)-Available on electronic scanning probes.  
            Request function: In multi-mode display, it is possible to select one image for full  
            screen display.    
            DMS (Integrated Data Management System)    
        1   Measurement and Analysis.    
            General measurements    
            On B-mode image    
            On M-mode image    
            On spectral Doppler    
            On B/D-mode image    
            On B(Fiow) mode image    
            Obstétrical measurements & calculations    
5   Ultrasound (3 probes)       Gynecological measurements & calculations    
        Peripheral vessels analysis    
               
            Urological measurements & calculations    
            Abdominal measurements    
            Report Functions    
            Data Communication Function    
            Preset Function    
            Characters and graphic displays    
            Probe connectors    
            Standard    
            Active connectors: 2 connectors    
            Dummy connector for rest: 1 connector    
            Viewing Monitor    
            15-inch diagonal color Liquid Crystal Display with brightness and contrast controls  
            Acoustic Power    
            0 to 100%, changeable    
            Environmental Requirements    
            Temperature: +10 to +40 degrees C    
            Relative Humidity: 30 to 75%    
            Power Requirements: 220V / 50 Hz.    
            Probes    
            General abdomen OB/GYN 2.5-6.0 MHz.    
            Small parts, PV (steered linear) 5.0-10.0 MHz.    
            Adult heart (harmonic echo) 2.1-3.8 MHz.    
6   Gastroscope (With halogen       Optical System :    
  light source)       Field of view: at least 100°°    
             
            Depth of field: at least 4 to 100 mm    

 

 

page 3 of 27

 

Equipment & Furniture Component

 

      Distal end : Outer diameter: not more than 10mm.
      Bending Section
      Range of tip bending: not less than Up 210°, Down 90°, Right 100°, Left 100°
      Insertion tube : Outer diameter: not more than 13.3mm
      Working length: not less than 1025mm.
      Length markings
      Instrument channel :
      Inner diameter: not less than 2.8mm.
      Light source
      Halogen technology shall provide a high light intensity of at least 150W.
      Variable light intensity.
      Fast and easy bulb change during operation : In case of a bulb failure immediate
    1 switch to the second bulb with the reverser shall be possible.Adjustment of
      brightness and power with buttons and switches on the front panel.
      Shall be connectable with conventional telescopes, fiberscope.
      Shall be adaptable to all endoscopic procedures
      Optionally , an automatic built-in swivel mechanism will bring the spare halogen
      bulb into working position.
      Technical Data : Halogen , 150 watts , Halogen, 2 pieces
      Manual switch , 2 x 150 W , 24 V
      Power Supply : 220 VAC 50 Hz .
      Standards
      CE;
      EC Marked
      US FDA;
      ISO certification
       
      Optical System :
      Field of view: at least 120°
      Depth of field: 5 –÷100 mm.
      Distal end : Outer diameter: not more than 13mm.
      Bending Section
    1 Range of tip bending: not less than Up 180°, Down 180°, Right 160°, Left 160°
      Insertion tube : Outer diameter: not more than 13.3mm
      Working length: not less than 1680mm.
      Instrument channel
      Inner diameter: not less than 3.2mm.
      Light Source
      Halogen technology shall provide a high light intensity of at least 150W .
      Variable light intensity
  Colonoscope   Fast and easy bulb change during operation : In case of a bulb failure immediate
7 (With halogen light   switch to the second bulb with the reverser shall be possible
   
  source)   Adjustment of brightness and power with buttons and switches on the front panel.
      Shall be connectable with conventional telescopes, fiberscope,
      Shall be adaptable to all endoscopic procedures
      Optionally , an automatic built-in swivel mechanism will bring the spare halogen
      bulb into working position.
      Technical Data : Halogen , 150 watts , Halogen, 2 pieces
      Manual switch , 2 x 150 W ,24 V
      Power Supply : 220 VAC 50Hz .
      Standards
      CE;
      EC Marked
      US FDA;
      ISO certification
       

 

 

page 4 of 27

 

    Equipment & Furniture Component  
         
      Recording ECG Leads: 12 standard Leads  
      Recording Channels: 3/1 user selectable  
      LCD display of ECG  
      Lead switching: manual and automatic.  
      Sensitivity, mm/mV: 5, 10, 20.  
      Calibration signal: automatic and manual.  
      Diagnostic frequency range 0.67–150 Hz or better  
      Filters for mains frequency, low frequency, muscle artefact, high frequency
      Recorder:  
      Recording method: thermal paper  
      Recording speed, mm/sec: 25/ 50 user selectable  
      Channels acquired simultaneously:1/ 3 user selectable  
      Channels printed simultaneously:1/ 3 user selectable  
      Other features:  
      Portable  
    3 Mains and internal rechargeable battery operation  
  Electrocardiograph 3 Battery operating time, minimum: 90 min.  
8    
  Power requirements: 220 V, 50 Hz  
channels    
       
      Accessories:  
      1. Patient cable  
      2. 6 chest limb electrodes  
      3. 4 limb electrodes  
      4. 4 strap electrodes  
      5.  
      6. 1 bottle ECG Gel  
      7. 2 rolls of paper or Z-Fold  
      8. Carry bag  
      Standards  
      CE;  
      EC Marked  
      US FDA;  
      ISO certification  
       
    3 Surgical Lighthead with five Projectors , Ceiling Suspended
      Light intensity (Ec) at least150 000 lx  
      Color rendering index (general) Ra 96  
      Color rendering index (red) R9 89  
      Light field size 15 – 30 cm.  
      Colour temperature 4200º Kelvin  
      Working range without focusing approximately 70 – 140 cm
      Power supply 220 VAC, 50 Hz  
      Adjustable light intensity  
      Power Supply , 220V/50 Hz.  
  Operating theatre light (5   Power Stabilizer shall be supplied  
9   This equipment shall be supplied with all necessary fixation gears to the Building
spotlights)  
    structure  
       
      Replacement Bulbs (5) and Sterilizable Handles (5) shall be supplied
      Standards  
      CE;  
      EC Marked  
      US FDA;  
      ISO certification  
      Accessories/Spares :  
      5 Bulbs per unit  
      2 Handles per unit  
      Operating table designed for carrying out treatments, dressing interventions and
  Operating Table   operations of general surgery and traumatology  
10 (Trauma/Multifunction,   All exposed metallic parts shall be made from stainless, acid proof steel.
  The table base shall be mobile and shall have central brakes
  electric/hydraulic driven  
    Back rest and leg rest inclination angle, kidney elevator, Trendelenburg and
      reverse-Trendelenburg positions and height adjustment of the table top shall be

 

 

page 5 of 27

 

Equipment & Furniture Component

    activated by electro – hydraulic system
    The table top shall be translucent for x-rays with 5 separate Sections
    Table top length approximately 2000 mm
    Table top width approximately 600mm
    Minimal height of the table 750mm
    Maximal height of the table 1100mm indicatively
    Trendelenburg at least 25
    Reverse Trendelenburg at least 25º
    Lateral tilt at least 18º
    Back rest inclination angle at least +55º to -25º
    Head rest inclination angle at least 45º up
    Head rest inclination angle at least 20º down
    Power requirements
    220 VAC ±10 % ,  50Hz
  1 Accessories:
    Anaesthetic Screen with clamp with telescopic tubes : 1
    Body restraint Strap with clamp : 1
    Padded Shoulder supports : 2
    Padded Leg support with swivel type clamp : 2
    Padded arm Rests 450 -500 mm long with two arm Clamps : 2
    Padded Lateral support with universal attachment Clamp : 2
    Padded rubber mattresses with Anti Microbial agent incorporated into all components
    that assists in Prohibiting growth of bacteria & fungi and easy to clean and maintain
    and of at least of 1” thickness : 2 X5 Sections
    Head rest , I.V drip Stands attachable to the table
    Environmental factors:
    The unit shall be capable of being stored continuously in ambient temperature of 0 –
    50deg C and relative humidity of 15-90%
    The unit shall be capable of operating continuously in ambient temperature of 10 -40
    deg C and relative humidity of 15-90%
    Accessory for traumatology are optional and shall be listed separately
    Standards
    CE;
    EC Marked
    US FDA;
    ISO certification
    Operating table for carrying out treatments, dressing interventions and operations
    of general surgery (3-4 surgeries daily).
    All exposed metallic parts shall be made from stainless, acid proof steel.
    The table base shall be mobile and shall have central brakes
    Back rest and leg rest inclination angle, Trendelenburg and reverse-
    Trendelenburg positions and height adjustment of the table top shall be activated
    by electro – hydraulic system
    The table top shall be translucent for x-rays with 5 separate Sections
    Table top length approximately 2000 mmTable top width approximately 600mm
    Minimal height of the table 750mm
    Maximal height of the table 1100mm indicatively
    Trendelenburg at least 25º
    Reverse Trendelenburg at least 25º
    Lateral tilt at least 18º
  Operating Table Back rest inclination angle at least +55º to -25º
11 (Multifunction, Head rest inclination angle at least 45º up
Head rest inclination angle at least 20º down
  electric/hydraulic driven)
   

 

Power requirements

 

220 VAC ±10 % , 50Hz Accessories:

 

Anaesthetic Screen with clamp with telescopic tubes : 1 Body restraint Strap with clamp : 1

  • Padded Shoulder supports : 2

Padded Leg support with swivel type clamp : 2

 

Padded arm Rests 450 -500 mm long with two arm Clamps : 2 Padded Lateral support with universal attachment Clamp : 2

 

Padded rubber mattresses with Anti Microbial agent incorporated into all components that assists in Prohibiting growth of bacteria & fungi and easy to clean and maintain and of at least of 1” thickness : 2 X5 Sections

 

Head rest , I.V drip Stands attachable to the table Environmental factors:

 

 

page 6 of 27

 

Equipment & Furniture Component

 

      The unit shall be capable of being stored continuously in ambient temperature of 0 –
      50deg C and relative humidity of 15-90%
      The unit shall be capable of operating continuously in ambient temperature of 10 -40
      deg C and relative humidity of 15-90%
      Standards
      CE;
      EC Marked
      US FDA;
      ISO certification
       
      Anesthesia Machine used for delivering anesthesia Agents to Patient during Surgery
      Procedures. The complete set-up shall also ventilate the Patient
      Patient monitoring system  ECG, EtCO2, Pulse Oximeter and airway pressure, NIBP,
      rectal/&skin Temperature.
      The full System shall be completed with :
      Anaesthesia gas delivery system, O2/N2O/Compressed Air .
      Circle absorber system.
      Precision vaporiser for halothane
      Anaesthesia ventilator
      Patient Circuit
      Anesthesia gas delivery system:
    3 Shall have provision for delivery of oxygen, nitrous oxide and medical air with
      pressure gauges.
      Oxygen and Nitrous oxide should be linked either mechanically or pneumatically to
      ensure a minimum of 25% oxygen delivery at all times to avoid delivery of hypoxic
      mixture.
      Shall have audio-visual oxygen Failure Warning System with Nitrous oxide cut off.
      Shall have back bar ISO pin type to attach Vaporizer easily.
      Flowmeter oxygen : 0÷2; 2÷10 litres/min, minimum range
      Flowmeter N2O : 0÷2; 2÷10 litres/min, minimum range with safety mechanism
      (hypoxic link) or similar
      Flowmeter , Medical Air 0÷10 litres/min.
      Shall be supplied with necessary Attachments for use of the breathing Circuits
      (Ruben , Bains, Jackson-Rees or Magill )
      Construction :
      Shall have top Shelf to keep monitors and a Tabletop with Drawers to keep
      anaesthetic drugs, equipments etc.
  Anesthesia trolley   Castor wheels should be durable , moisture resistant and antistatic
    The Anaesthesia machine frame shall be made of rust proof material/Stainless steel
12 (anesthesia machine with  
  or sheet Steel protected with Epoxy-Paint
  ventilator and monitor)  
    Standard Circle Absorber System
     
      Shall have adjustable pressure limiting Valve, breathing circuit pressure measuring
      device.
      Shall have a Bag/Ventilator selecting Valve integrated onto the Absorber.
      Facility to attach oxygen sensor
      Shall have CO2 absorbent Chamber Canister (Soda Lime )
      Precision Vaporizers (Temperature, Pressure and Flow compensated) for Halothane
      Should be easy to mount and dismount from the back bar.
      Vaporisers shall have ISO pin type (Selectatec) Mounting and Vaporiser interlocking
      facility
      Shall have a standard filling port with keyed filling device
      Should be designed for transport with liquid in Vaporiser Chamber with protection
      against tipping and shaking
      Maintenance free Vaporiser
      Ventilator
      Shall be a bag in bottle anaesthesia Ventilator with standing (ascending/Piston)
      Bellows
      Shall be supplied with adult and paediatric Bellows
      Shall be able to set Tidal Volume, Respiratory rate and I:E ratio
      Ventilator should have audible alarms for ventilator failure, low oxygen supply
      Pressure, inadequate volume delivery, disconnection alarm, and power supply failure.
      Ventilation mode: CMV-SV – manual , Minute volume: 2 – 30 l/Mn
      Tidal volume: approx.50 – 1400ml , Breathing frequency: approx.6-60 times/min

 

 

page 7 of 27

 

Equipment & Furniture Component

Flow rate range: 5 – 60 l/min

High pressure relief valve

Power supply 220VAC, 50Hz

 

Accessories :

 

Accessories required for Ventilator :

Resuscitators , Adult / Child : 1 Each

Set of face masks , 3 Sizes : 2 Each

Corrugated hoses and tubing set for Pipeline O2 : 2 Sets

 

Corrugated hoses and tubing set for Pipeline Air : 2 Sets

 

Accessories required for Anaesthesia Machine :

Hose assembly for piped Oxygen supply : 2 Sets

 

Hose assembly for piped Nitrous Oxide supply : 2 Sets

Hose assembly for piped Air supply : 2 Sets

 

Regulator and manometer for the pipe pressure hose ( 60 psig) : 1 Set

 

Connection from ventilator to anaesthetic apparatus , autoclavable/reusable : 1 Set

Test bag, 1 litre : 2

Peak Flow Meter : 1

Power cord with grounding wire : 1

 

Dust cover : 1

Bellows assembly, adult / Children : 2 Each

Patient Circuit , Magill Type , Complete , Adult / Children . . Shall be Reusable : 2

Each

Spirometer (Peak Flow Meter for Anaesthesia procedures)

Consumables for anaesthetic unit for 2 years operation:

Battery cell (as appropriate)

Soda lime for circle absorber (5Kg/pack) : 10 Packs for 1 unit

 

Standards

 

CE;

 

EC Marked

US FDA;

ISO certification

Syringe Driver for Parenteral Infusions , in Wards, OP, ICU’s and during Transportation/Transfer

Single Channel. Shall accept Up to 60cc Syringes , Non-Captive

 

Self checking of Electronics with audible call back on locked-out Controls Tamper and fluid resistant Design . ABS Casing (Impact-Resistant )

 

Shall be fitted with IV Rod, Adjustable Height with 4 Hooks.Mobile on 5 Star Stand with Castors

 

Internal rechargeable battery

 

Rate: 0.1 to 99.9 ml ( in 0.1 ml steps) indicatively Total volume infused: 0.1 to 999.9 ml

Precision : +/-3%.Stipulated Syringes

Occlusion release shall give minimum Bolus effect

 

Syringes : 10 Types at least of LuerLock Type 10/60mL.Syringes,+ Specific Choice

Pause infusion facility

Alarms/Safety

 

13 Syringe pump Infusion nearly complete Infusion complete Occlusion

 

System malfunction Syringe unlocked Plunger disengaged Low battery

 

Protection against Leakage Current:Type CF.Equipment

Protection against Electric Shocks:Class II.Equipment

Safety features: Error coding (event log)

Rate change through Audible Stop Switch

Audible Call back on locked-Out-Controls

  • Automatic detection of Syringe Volume

 

Minimum bolus on Occlusion Release at any selected Rate Communication : RS232C Interface

 

Electrical:

Electrical: Class II (double insulated)

 

 

page 8 of 27

 

Equipment & Furniture Component

 

 

      Bat.Autonomy : 8 Hrs.Minimum
      Power supply: 220 VAC.50.60Hz
      External transformer . Stabilizer
      Environmental factors
      Shall meet IEC-60601-1-2: 2001(Or Equivalent BIS)
      General Requirements of Safety for Electromagnetic Compatibility or should
      comply with 89/366/EEC; EMCdirective.
      The unit shall be capable of being stored continuously in ambient
      temperature of 0 -50deg C and relative humidity of 15-90%
      The unit shall be capable of operating continuously in ambient temperature of
      10 -40deg C and relative humidity of 15-90%
      Accessories
      Coupler between 2 units
      Mobile floor stand
      IV stand clamp
      Rail clamp
      Standards :
      IEC601.1/IP33
      Safety: IEC 601 / 1. CF.
      Compliance with EN 60601.1 and PrEN 60601-1-24.
      CE 0459 marking in compliance with EEC 93/42 Medical Products directive.
      Comply with  EN60601-1-2, EN60601-2-24
      EN ISO 9001:2000, EN ISO 13485:2000; EN 46001:1996
       
      HF electrosurgical unit shall be used to execute monopolar and bipolar surgery in
      many fields of application where high precision and reliability are essential
    3 Outputs of cut, coagulate and blend
      Maximum output 300 W for monopolar cut
      Activation :
      Double pedal switch which may be used for the monopolar and bipolar functions .
      Hand-switch handle
      Bipolar electrode with pedal switch or with automatic Start/Stop system ( for
      coagulation only)
      Control
      The Unit shall stop automatically in case of internal error which shall be identified on
      Display and with audible alarm
      Memorization : User shall be able to use at least 4 working programs
      Safety :
      Neutral plate safety circuit shall control connections and contacts of Neutral Plate
      with Tissues:
      Defective Contact shall be notified with visual Alarm and immediate reducing of
      power
      Output circuit : floating – protected against defibrillator interferences . Shall have HF
  Electrosurgical   leakages less than 150mA through each electrode
14   Power Supply : 220VAC, 50Hz
Unit (Monopolar-bipolar)  
    Cooling: convection without fan
     
      Accessories :
      Single-use two button Handle , Autoclavable
      Handle shall be provided with a 3 Pins socket that shall fit majority of bipolar
      electrosurgical units
      Operative Foot-switch (usable as alternative to handle)
      Reusable neutral Plate
      Kit of 10 short autoclavable Electrodes
      Electrode wire – straight
      Electrode wire – angled 45°
      Electrode bend – 0 4 – straight
      Electrode bend – 0 8 – straight
      Electrode hook – angled 45°
      Electrode wire 1 mm.- angled 45°
      Electrode straight slip-knot – straight
      Electrode angled slip-knot – straight
      Electrode ball point – 0 3 mm. – angled 45°
      Electrode ball point – 0 3 mm. – straight
      Bipolar Forceps : 6
      Forceps shall be provided with a standard European connection
      Straight Forceps – 18 cm. (7″)

 

 

page 9 of 27

 

Equipment & Furniture Component

 

      Curved Forceps – 18 cm. (7″)  
      Curved Forceps – 20 cm. (7 3/4″)  
      Bayonet Forceps – 18 cm. (7″)  
      Bayonet Forceps – 20 cm. (7 3/4″)  
      Straight Forceps – 20 cm. (7 3/4″)  
      Cable, Bipolar  
      Adaptor, Bipolar Cable  
      User Manual  
      Standards  
      CE;    
      EC Marked  
      US FDA;  
      ISO certification  
       
      Suction Unit for Major Surgery Procedures . Mains-powered , mobile on 4 antistatic
      Castors, ABS Casing and 2 graduated Canisters of 2,000ml each made of
      Polycarbonate autoclavable at 121°C and disposable suction bags  
      Shall require no maintenance nor lubrication  
      Oil-free pump, maximum suction of at least 500 mm Hg  
      Free flow rate at least 25 l/min  
      Main Switch with Pilot Lamp . Fuses  
      Pedal Action  
      Shall be equipped with a protective thermal cut-out relay.  
      Shall be equipped with motor-protection cap that totally prevents aspirated liquids or
      secretions from reaching and damaging the vacuum pump  
      Suction command with continuous adjustment , Vacuometer  
      2X2,000ml Canisters with airproof screwing-cap with independent Overflow devices
      . Fast Connectors and silicone Tubing  
      Power Supply : 220VAC.50Hz.  
      Ventilation Fan for overheating  
15 Suction Pump 2 Sound level: Shall be not more than 55 dBA  
      Accessories  
      Silicone Tubing , sterilizable  
      Transparent Cannula Holder, Sterilizable  
      Anti-Bacterial Filters ( 4 )  
      Set of 4 canulaes with Holder : Yankhauer , Soft Universal Yankhauer
      Diameter : 8.0/6.0mm with anti-sticking Lumen and High Suction Lumen
      Universal Soft Canulaes diameter : 6.0/4.0mm  
      Frazier Canulaes (Fergusson) diameter :1,5/2.0/3.0/4.0mm  
      Jackson Canulaes : 35×3 /45X3/25X4/35X4/45X4/55X4/60X4cm  
      Standards  
      CE;    
      EC Marked  
      US FDA;  
      ISO certification  
      1 Operating Scissors, 170 mm, straight, blunt-blunt points 1
      2 Operating Scissors, 140 mm, straight, sharp-blunt points 1
      3 Operating Scissors, 140 mm, curved, sharp-blunt points 2
      4 Dissecting Scissors, curved, for deep surgery, 230 mm 2
      5 Operating Scissors, 140 mm, curved, blunt-blunt points 4
      6 Operating Scissors, curved, blunt-blunt points, 170 mm 2
      7 Operating Scissors, 140 mm, curved, hardened, blunt-blunt 1
      points
         
16 Operating instrument set   8 Sharp Tips Scissors, curved, 160 mm 1
    2 9 Dressing Forceps, standard pattern ПА 150х2,5 6
      10 Dressing Forceps, standard pattern ПА 200х2,5 4
      11 Surgical Forceps, 150×2,5 8
      12 Surgical Forceps, 1 x 2, 200×2.5 mm 4
      13 Abdominal Depressor, 285 mm, blade 70 mm wide 1
      14 Tissue Forceps, 265 mm, straight 5
      15 Tissue Forceps, 256 mm, curved 5
           

 

 

page 10 of 27

 

  HSMP Armenia Equipment & Furniture Component     HPIU: HH
           
      16 Hemostatic Forceps, curved, atraumatic (160 mm) 6
      17 Hemostatic Forceps, curved, 1×2 teeth, serrated, size 2, 6
      (160mm)      
               
      18 Hemostatic Forceps, curved, serrated, size 2 , (196 mm) 6
      19 Hemostatic Forceps, curved, serrated, (240 mm)   6
      22 Hemostatic Forceps, straight, 1×2 teeth, serrated, size 1 , 3
      (150 mm)      
               
      23 Hemostatic Forceps, curved, 1×2 teeth, serrated, size 1, 6
      (150mm)      
               
      24 Hemostatic Forceps, curved, 1×2 teeth, serrated, size 1, 6
      (200mm)      
               
      25 Hemostatic Forceps for deep surgery, child size, size 2, 6
      (225mm)      
               
      26 Rake Retractor, 2 sharp prongs, (200 mm)   2
      27 Rake Retractor, 3 sharp prongs, (200 mm)   2
      28 Rake Retractor, 200 mm, 4 sharp prongs, style 2   2
      29 Rake Retractor, 2 blunt prongs, 200 mm   3
      30 Rake Retractor, 220 mm, 4 blunt prongs, style 2   4
      31 Rake Retractor, 6 sharp prongs, (200 mm)   2
      32 Abdominal Retractor       2
      33 Needle Holder,( 200 mm)       2
      34 Needle Holder, ( 250 mm)       2
      35 Vascular Needle Holder , 160 mm     2
      36 Vascular Needle Holder, 200 mm       2
      37 Intestinal Forceps, non-crushing, curved, child size, (192mm) 2
      38 Intestinal Forceps, non-crushing, curved, adult size, (235mm) 2
      39 Tongue Spatula, straight, 20х16mm, (180 mm)   2
      40 MОSQUITO Hemostatic Forceps, serrated, curved on edge, 6
      61/10″ (154 mm)      
      41 Intestinal Forceps, non-crushing, straignt, adult size, 2
      (240mm)      
               
      42 Abdominal Retractor (60 mm)       2
      43 Abdominal Retractor (100 mm)       2
      44 Liver Speculum (100 mm)       2
             
       
      Suction Unit for Emergency procedures , fitted with single 1000ml capacity
      Polycarbonate Canister , mobile on 4 Castors and carrying handle  
      The Unit shall enable operation from either mains supply, or independently from an
      internal 12V rechargeable battery        
      The unit shall recharge when plugged into the mains    
      An indicator on the casing shall indicate the status of the battery . A Vacuometer
      shall indicate the suction value        
      Maintenance free diaphragm pump        
      Overflow safety device        
      The Unit shall be capable to generate 500mm Hg at least  
      Free flow rate at least 25 litres per min        
      Shall be fitted with easily replaceable anti-bacterial filter with 5 spare
    1 Shall provide over 30 minutes of full vacuum from a fully charged battery
17 Suction Pump   Fitted with rechargeable battery        
      Operating mains voltage: 220 V – 50 Hz        
      Recharging is by trickle charge and shall take eight hours maximum to restore full
      charge        
      Accessories :        
      Hand piece with regulator and appropriate probes shall be supplied as well as an
      additional Canister        
      Standards        
      CE;          
      EC Marked        
      US FDA;        
      ISO certification        
       
      The following specifications define a portable, lightweight ; battery or accumulator
18 Defibrillator   operated automated external to treat Patients requiring basic Life Support
  Weight of complete unit, excluding batteries or accumulator , shall permit easy
     
    2 Transport . The overall Dimensions shall enable easy storage and portability

 

 

page 11 of 27

 

Equipment & Furniture Component

 

      The Casing shall be constructed to withstand the standard operating Conditions
      in a Physician’s office and/or Hospital. The device must also with stand the
      harsh operating conditions associated with Ambulance use. The unit shall be safe to
      use both for the operator and the patient.
      Defibrillator
      The device shall utilize Biphasic waveform technology as the primary energy delivery
      option and shall include automatic and manual modes
      The biphasic waveform shall allow for escalating energies up to at least 200 Joules
      Safety: Internal discharge of energy if no discharge is effected.
      Internal self-testing and fault recognition.
      Required power 200-240V 50/60 Hz
      Weight not more 10kg
      Accessories :
      Full Standard Accessories Set-Up shall be supplied
      Standards
      CE;
      EC Marked
      US FDA;
      ISO certification
       
      Intensive care ventilator, for use on Pediatric/Infant- Adult patients
      With Graphics Display Monitor.
      Breath Types: VC, PC, VTPC
      Modes ( in each breath Type): A/CMV, SIMV, SPONT,
      Spontaneous Breath Choices: PS, VTPC.
      Features:
      Inspiratory time at least 0-3 sec with optional pause
      Inspiratory flow at least 3-140 l/min
      Pressure support: at least 0 to 45cm H2O/mbar
      Tidal Volume: at least 10-800ml
      Resp.Rate: at least 6-150b/min
      Pressure Limit: 0 to 80cmH2O/mbar
      I:E Rate: 1:4 to 4:1
      Trigger pressure or flow
      FiO2: 0.21 to 1.00
      PEEP/CPAP: 0 to 45 cmH2O/mbar
    2 Leak Compensation
      Standby Condition: Allows setting to be preset and Circuit Check tests to be
      performed prior to starting Ventilation
      Open Exhalation Valve (BPRV): On/Off for Biphasic Pressure Release
      Ventilation
      Event History Log: Records 1000 events; alarm & settings
19 Ventilator ICU   Compliance Compensation: On/Off
  Graphical display of respiratory parameters
     
      Must be mounted over the Mobile Medical Air Compressor.
      Power requirements
      220 VAC ±10%,  50Hz.
      Internal rechargeable Battery provides an minimum. 60 minutes for complete
      ventilator function.
      Accessories :
      Medical Air Compressor
      Humidifier
      1 Tube set, adults
      1 Tube set, Children
      1 Humidifier chamber
      O2-connecting Tube 5.0 m
      1 Air-connecting Tube 5.0 m
      Humidifier : Heating power with safety thermostat and indication of operating status
      1 Circuit , Adults with patient Tubes , water traps, Y-Piece,mask elbow, catheter
      connector
      1 Circuit , Children with patient Tubes, water traps, Y-Piece,mask elbow, catheter
      connector
      1 Carriage
      Articulated Bracket for rail
      O2-cell without housing

 

 

page 12 of 27

 

Equipment & Furniture Component

 

      Sensors 5 pcs set
      Temperature Sensor
      Humidifier
      Bag- test-lung
      Standards
      CE;
      EC Marked
      US FDA;
      ISO certification
       
      Real-time SpO2%
      Oxygen Saturation Range (%SpO2)
      Measurement range at least 70-99%
      Accuracy (%SpO2) ±2% at100%~80%, ±3% at 80%~70%
      Real-time heart rate (BPM)
      Pulse
      Measurement range 30-250 bpm
      Accuracy ± 1 bpm at 30~250 bpm
      Display
      SpO2,Pulse rate, Pulse signal wave, Max/Min setup (SpO2% / BPM)
      Power Requirements internal rechargeable battery providing an operating time of at
      least 8 hours
      Compact design for easy transportability
20 Pulsoximeter    
    2 Accessories:
      3 rolls of paper
      Operation manual
      1 adult reusable finger Sensor
      1 Neonate Sensor
      Standards
      CE;
      EC Marked
      US FDA;
      ISO certification
       
      Patient monitor that shall draw Cardiograms, measures Blood Pressure (invasive or
    7 not-invasive type), Respiration values, Temperatures at two spots, Oxygen Saturation
      in arterial blood ( SpO2 ), and Pulse rate ( HR ) .
      May be used to Monitor Patients on the transportation by using batteries, and the
      measured information on patient’s condition shall be displayed in wave forms along
      with numerical values.
      Display/Screen
      Display : Colour TFT display , screen diagonal Curves and measured values : not less
      than 25.00cm . Up to 6 Curves as well as numerical fields
      Brightness: can be manually adjusted
      Sweep Speed : Shall adjuste to 6.25, 12.5, 25 and 50 mm/s
      Display pixels: at least 1024×768
      Monitor Interfaces
      R-SYNC: for defibrillator , Each R-wave shall be clearly identified , Staff call:
21 Patient Monitor, ICU   Remote alarm via alarm adapter
  RS-232: as computer interface
     
      Mains/Battery Operation
      Mains operation: shall be via built-in Power Supply Battery operation: via integrated
      battery or Charging console . Battery operating time: approx. up to 4 hrs at least
      Power Supply : 220 V AC ±10%, automatic change-over Power failure buffer : for
      trend values,
      System Expansion
      Software updates via interfaces
      Monitoring parameters shall include : ECG, RESP, IBP, NIBP, SPO2, PULSE RATE
      , DUAL-TEMP
      S-T segment synchronized detection and analysis
      Graphic and tabular trends of all parameters for at least 96 hours
      20-Mn minimum ECG real time waveform and parameters storage and review
      Shall be able to be connected with a Central Monitor Station

 

 

page 13 of 27

 

  HSMP Armenia Equipment & Furniture Component HPIU: HH
       
      Inner high-definition thermal dot matrix recorder optional, which can output
      waveforms and characters  
      ECG : Input : Whole-lead ECG cable, standard AAMI cable connector
      Sweep speed: 12.5, 25, 50mm/s 0.5~40Hz(for monitor or operation)
      Calibration signal: 1mV , Protection: Against electrosurgical interference and
      defibrillation  
      Heart Rate: Measuring range: 30~250bpm , Alarm mode: Audible and
      visualAlarm  
      Respiraton Measurement  Range: at least 0~60 bpm  
      Oxygen Saturation(SPO2) Meas. Range: at least 70~100%
      NIBP : Method: Oscillometric , Mode: Manual, Auto , Resolution: 1 mmHg
      Shall have Over-pressure protection: (Adult 300mmHg ; Child 220mmHg
      Temperature : Meas. Range: at least 20~50C , Resolution: 0.1 C , Accuracy: +/- 0.2 C
      Operation Environment : Temperature: 0~45C , Humidity: 30~85%(non-condensing)
      Accessories  
      Electrode cable for disposable electrode (3 leads)  
      NIBP cuff for adults  
      NIBP cuff for children  
      Hose for NIBP cuff  
      Respiration Sensor  
      Power cord  
      Disposable electrode (5 pcs)  
      Temperature sensor – Adult (rectal)  
      SpO2 sensor reusable  
      Console / Bedside Holding Device  
      Standards  
      CE;  
      EC Marked  
      US FDA;  
      ISO certification  
       
      Patient monitor that shall draw Cardiograms, measures Blood Pressure (invasive or
      not-invasive type), Respiration values, Temperatures at two spots, Oxygen Saturation
      in arterial blood ( SpO2 ), and Pulse rate ( HR ) , End Tidal CO2 (etCO2) .
    2 May be used to Monitor Patients on the transportation by using batteries, and the
    measured information on patient’s condition shall be displayed in wave forms along
     
      with numerical values.  
      Display/Screen  
      Display : Colour TFT display , screen diagonal Curves and measured values : not less
      than 25.00cm . Up to 6 Curves as well as numerical fields  
      Brightness: can be manually adjusted  
      Sweep Speed : Shall adjuste to 6.25, 12.5, 25 and 50 mm/s  
      Display pixels: at least 1024×768  
      Monitor Interfaces  
      R-SYNC: for defibrillator , Each R-wave shall be clearly identified , Staff call:
22 Patient Monitor, anesthesia   Remote alarm via alarm adapter  
  RS-232: as computer interface  
       
      Mains/Battery Operation  
      Mains operation: shall be via built-in Power Supply Battery operation: via integrated
      battery  
      or Charging console . Battery operating time: approx. up to 4 hrs at least
      Power Supply : 230 V AC ±10%, automatic change-over Power failure buffer : for
      trend values,  
      System Expansion  
      Software updates via interfaces  
      Monitoring parameters shall include : ECG, RESP, NIBP, SPO2, PULSE RATE ,
      DUAL-TEMP and etCO2  
      S-T segment synchronized detection and analysis  
      Graphic and tabular trends of all parameters for at least 96 hours
      20-Mn minimum ECG real time waveform and parameters storage and review

 

 

page 14 of 27

 

    Equipment & Furniture Component  
       
      Shall be able to be connected with a Central Monitor Station
      Inner high-definition thermal dot matrix recorder optional, which can output
      waveforms and characters  
      ECG : Input : Whole-lead ECG cable, standard AAMI cable connector
      Sweep speed: 12.5, 25, 50mm/s 0.5~40Hz(for monitor or operation)
      Calibration signal: 1mV , Protection: Against electrosurgical interference and
      defibrillation  
      Heart Rate: Measuring range: 30~250bpm , Alarm mode: Audible and visual Alarm
      Measurement Range: at least 0~60 bpm  
      Oxygen Saturation(SPO2) Meas. Range: at least 70~100%
      NIBP : Method: Oscillometric , Mode: Manual, Auto , Resolution: 1 mmHg
      Shall have Over-pressure protection: (Adult 300mmHg ; Child 220mmHg
      Temperature : Meas. Range: at least 20~50C , Resolution: 0.1 C , Accuracy: +/- 0.2 C
      End Tidal CO2 (etCO2) Module , Non Dispersive Infrared Spectroscopy Method
      Mainstream or sidestream  
      Displayed Data : Waveform labels and annotations , etCO2, inC02 and respiration
      rate values  
      etCO2 Alarm Limits  
      Operation Environment : Temperature: 0~45C , Humidity: 30~85%(non-condensing)
      Accessories  
      Electrode cable for disposable electrode (3 leads)  
      NIBP cuff for adults  
      NIBP cuff for children  
      Hose for NIBP cuff  
      Respiration Sensor  
      Power cord  
      Disposable electrode (5 pcs)  
      Temperature sensor – Adult (rectal)  
      SpO2 sensor reusable  
      EtCO2 Mainstream or Sidestream accessories , Transducer , Lines , Filters
      Console / Bedside Holding Device  
      Standards  
      Classification: CF, defibrillation-proof,  
      CE;  
      EC Marked  
      US FDA;  
      ISO certification  
      Compact and easy to transport oxygen concentrator ( Mobile on Castors )
      The Oxygen Concentrator shall feature a dual-head Compressor. One head
      producing. Capacity : 1 to 5 l/Mn of O2 at 90% minimum at maximum flow
      Pressure-compensated Flowmeter shall permit use of long canula while maintaining
      accurate flow setting.  
      Safety Alarms: Audible and visual : Power Failure , Restricted Flow , , Low O2
      Pressure-relief Valve and thermal protection of the Compressor
      Double-insulated Unit , Two-prong plug  
      Flame-retardant Cabinet  
      Sound Level (ANSI): 50 dB average  
23 Oxygen Concentrator   Fixed humidifer Port and Recess shall prevent bottle and connector breakage
     
    1 Power Requirements: 220 VAC,50 Hz  
       
      Accessories :  
      O2 Tubing , 3 Face Masks ( Adult , Infant , New Born ), Humidifier ,
      Set of Filters  
      Standards  
      CE;  
      EC Marked  
      US FDA;  

 

 

page 15 of 27

 

    Equipment & Furniture Component  
         
      ISO certification  
      Manufactured from chrome plated steel tube, three sections with independent
      manual controls and Trendelenburg position.  
24 Couch, Gynecology 4 Top is upholstered with washable plastic material and flame retardant.
      Complete with padded Goepel leg-holders and S.S. bowl.  
      Knock down construction.  
      Manufactured from reinforced epoxy-painted steel tube  
      Three sections with independent manual controls.  
      Trendelenburg position.  
25 Delivery Bed   Third section extensible.  
    2 Top is upholstered with washable plastic material flame retardant.
      Complete with S.S. tray, and two adjustable knee crutches, padded shoulder rest,
      handgrips, leg-holders.  
      Lamp , Examination . Mobile on 5 Castors , 5 star-shaped Heavy and stable Stand .
      Halogen Type  
      Height of LampHead shall be adjustable  
      Light Intensity shall be at least 15000 Lux with a Field radius of 17cm Minimum
      Color Intensity : 4300 °K. Single Reflector of 150mm Minimum
      Power rating : 1X40W Minimum . 220VAC/50Hz . Transformer , 24V DC shall be
26 Examination Lamp 3 included  
  Accessories /Spares : 5 Halogen Bulbs  
       
      Standards  
      CE;  
      EC Marked  
      US FDA;  
      ISO certification  
27 Uterine Aspiration Set 1 Stainless steel with handle for suction regulation.  
  Cannule ø mm. 6/8/10/12/14  
       
       
      CTG or Fetal monitor for recording and analyzing the fetal heart rate (FHR) on beat-
      to-beat basis and the uterine contraction with Doppler -US transducer and taco
      transducer.  
      Portable handheld CTG monitor. Compact and battery operated, shall allow Patients
      to be continuously monitored anywhere . Color display shall show the CTG data
    1 presented in standard format for easy interpretation.  
      Fetal signals shall be easily located using the wide angle transducers with
      contractions stress being monitored by an external pressure transducer. Audio signals
      of each fetal heart are available on selection by the user through the built in Audio
      system.  
      The user will have control over the audio volume, audio fetal channel, chart recorder
      speed, toco zero, and event mark via the units front panel push buttons.
      Directional and watertight Ultrasound and TocoGraph with color-coded transducers
      for Single and Twins  
      Transducer: Multi element wide angle  
      Range: 30 to 240 bpm  
      Power output: < 20mW/cm2  
      Toco  
28 Fetal Monitor   Transducer: Real Time uterine Activity (contractions) trace on standard scale format
  Digital autocorrelation shall provide accurate and reliable FHR traces
     
      A pre-cut, non-fray elasticised belt with Buckle shall enable easy Transducer
      positioning for more accurate traces in difficult monitoring situations
      Clinical Event marker & automated foetal movement detection . Trace annotation
      Comprehensive patient database  
      Shall store over 500 patient records (30 minutes traces) in internal memory of FHR,
      Toco , Fœtal Movements with unlimited data storage on plug-in PCMCIA memory
      Cards  
      Computer interface 3 wire RS232  
      Thermal printer for common CTG paper  
      Speeds: 1,2,3, cm/min  
      Paper: z-fold pre-printed chart scale  
      The grid and the FHR/UA trace shall be printed simultaneously , ensuring printing
      accuracy and preventing potential errors in interpretation due to paper drifting or pen

 

 

page 16 of 27

 

    Equipment & Furniture Component  
         
      jitter  
      Power Supply :220 V AC and DC battery operated  
      Accessories:  
      1. Belts : 1 Original + 1 Spare Set  
      2. Buckles  
      3. IUP Starter Kit  
      4. Multicrystal transducer  
      5. Toco Transducers  
      6. Event marker Consumables:  
      7. CTG printer paper : 20 Packs  
      8. Fetal scalp electrodes  
      9. Ultrasound gel  
      10. IUP catheters  
      Standards  
      CE;  
      EC Marked  
      US FDA;  
      ISO certification  
      Capacity-0-15 kg  
29 Scale for newborn children 3 Graduation-10 gr.  
Precision-+/- 10 gr.  
   
       
      Washable surface  
         
      Body with 6V 20W Halogen for 220V, 50 Hz.  
      Quadruple revolving nosepiece.  
      Sturdy,base with supportive rubber feet.  
      With low position, coaxial coarse and calibrated fine focus control, with graduation
      reading 2 microns per division.  
      Total focusing 40 mm.  
      Condenser adjustable.  
30 Microscope Binocular 1 Built-in Power supply , halogen bulb adjustable.  
Double layer mechanical specimen stage.  
       
      Dust cover, blue filter, immersion oil and instruction manual included.
      2 pcs. 6V 20W halogen bulbs included.  
      Binocular Head, 30° inclined, 360° rotatable, Interpupillary Distance 55 – 75 mm
      Objectives:  
      ACHRO 4x/0.10  
      ACHRO 10x/0.2  
      ACHRO 40x/0.65, spring loaded  
      ACHRO 100x/1.25, Oil, spring loaded  
       
      Microscope with metal base with 360° rotatable monocular tube, 45° inclined
      3-fold objective nosepiece  
      separated coarse and fine focus  
      Stage focus control (protection of sample!)  
      Objectives:  
      ACHRO 4x/0.10  
      ACHRO 10x/0.25  
      ACHRO 40x/0.65 spring  
31 Microscope Monocular 1 Objective 100x/1.25 oil,  
      attachable mechanical stage, etc.  
      Abbé condenser n.A. 1.25 with iris diaphragm, variable at height
      Stage with stage clips  
      Blue filter.  
      Built-in illumination 6V/20 Watt Halogen, adjustable  
      Daylight mirror for outdoor-use  
      Dust cover.  
      Packed in styrofoam case  
      Power Requirements: 220V / 50 Hz.  

 

 

page 17 of 27

 

Equipment & Furniture Component

 

To provide accurate and precise hematology results with WBC 3 parts differential. Principle method of detection: volumetric impedance, electrical impedance, laser flow cytometry, hydrodynamic focusing or equivalent.

 

It should be able to provide precise measurement of

  • White Blood Cells count (WBC): Precision (CV, coefficient of variation) < 3.5%
  • Red Blood Cells (RBC) : Precision (CV) < 2%
  • Platelets (PLT) : Precision (CV) <6.0%, linearity range: 10 – 950×109/L
  • Hemoglobin (Hgb)
  • Hematocrit (HCT)
  • Mean corpuscular volume (MCV)

 

  • Mean corpuscular hemoglobin (MCH)
  • Mean corpuscular hemoglobin concentration (MCHC)

 

  • Red blood cell distribution width coefficient of variation (RDW-CD) 1 10) Mean platelet volume (MPV)

 

  • Lymphocyte (L% & L#)
  • Monocyte (M% & M#)
  • Granulocyte (G% & G#)
  • Others

 

Please indicate all parameters included in the offer here Minimum WBC 3 parts differential

Minimum 3 histograms

 

32 Shall have dual sampling mode of both whole blood as well as capillary vascular Analyzer Hematology Efficient work flow control with patient selective testing

 

Shall be an open system to accept generic reagents and/ or local made reagents All reagents shall be cyanide free

With build-in printer

 

With quality control function and reporting

With programmable automatic cleaning and maintenance

 

Accept samples with any anticoagulant including EDTA, heparin and citrate acid Data storage of approximately 300 samples.

 

Approximately 60 samples/hour With operation menu, reporting

 

Possible with interfacing capabilities to external PC, Built-in Printer, LIS software or Network

 

The system offered shall be designed to operate normally under the conditions of the purchaser’s country. The conditions include Power Supply, Climate, Temperature, Humidity, etc.

 

Please declare in detail compliance of this item offered with any relevant quality and safety standards.

RS232

 

Power Supply : 220VAC.50Hz. Accessories :

Start-up Reagents as per Manufacturer

      Compact semi-automated urinalysis instrument with a throughput of 50 tests/h  
      Shall have individual setting of concentration ranges, grades and units  
      Automatic urine colour determination  
      Memory capacity of 150 results at least  
      Interface ports for connection to PC , bar-code reader and external printer  
      Results based upon test strips  
      Shall have long period calibration stability  
      Automatic compensation of intrinsic urine colour avoiding false positive results  
    1 Power Supply :  
  33 Urinalysis System 220VAC.50Hz .  
  Accessories:  
       
      Strips , 1000 pcs  
      Calibration Testing Trips  
      Standards  
      CE;  
      EC Marked  
      US FDA;  
      ISO certification  
         
  34 Chemistry Analyzer At least 100 tests/hr  
  Semi- Automatic Including end point and kinetic measurement  
     
      Capable of analysing serum, plasma, urine and CSF  
         

 

 

page 18 of 27

 

Equipment & Furniture Component

Automatic calibration

Indication of abnormal values

Computer interface

Analysis of Albumin, ALP, ALT, AST, BUN, Ca++, Cholesterol, CK, Cl-, CO2,

Creatinine, Direct and total bilirubin, Glucose, Inorganic phosphorus, Iron, K+,

 

LDH, Mg, Na+, Total protein, Triglycerides, Urea, Uric acid and others

 

    1      
      Coagulometer with 2 independent operating channels , Semi automatic operation  
      Tests performed: PT, APTT, Fibrinogen, cofactors II, V, VII, VIII, IX, X  
    1 Optical detection with magnetic stirring used to detect clots  
      Temperature regulated at 37ºC ± 0.1ºC  
      Automated test calculation following addition of reagent, automatic compensation  
      for blanks    
      Straight forward input of test parameters, incubation period, tolerance for test  
      performed in double, reagent data (ISI, unit)  
      Automatic calibration/self test system  
      Capability to store calibration curves, presenting results in different units (second,  
      %, ratio, g/l or mg/ml.  
      Capacity PT: 60 in duplicate/per hour, APTT 45 in duplicate/per hour  
      LCD/LED display  
      Incubator capacity:  10 tubes  
      RS 232 interface for transferring results to PC or printer  
      Power supply: 240VAC, 50Hz .  
      Accessories :    
      Thermoprinter    
35 Coagulometer   Starter kit consisting of 100 pcs of cuvettes with magnetic stirs  
      Consumables:    
      PT tests    
      APTT tests    
      Fibrinogen tests  
      Printer paper 20 meter rolls  
      Reagent cuvettes, 100 pcs  
      Magnetic stirrers, 10 pcs  
      Standards    
      CE;    
      EC Marked    
      US FDA;    
      ISO certification  
         
      Precise Temperature Control up to 99 ºC  
      Model with Capacities : 6 L  
      Seamless Stainless Steel Bath  
36 Water Bath 1 Range Ambient +5ºC to 99ºC  
      Temperature Accuracy ± 0.5ºC  
      Uniformity ± 3.0ºC  
      Power Requirements: 220VAC / 50 Hz.  
         
      Productions rate: not less than 4 L/Hr  
      Standard equipment includes low water cut-off protecting heating elements from  
37 Water Still 2 burnout when water supply is interrupted  
  Drainage Tap    
         
      Power Requirements: 220V / 50 Hz.  
         
      Centrifuges are required in the Laboratory to separate various components of Blood  
      for analysis.    
      Operational Requirements :  
38 Centrifuge   Table top version  
      Technical Specifications:  
      Tube Capacity :No. 24 – 36 : Size 5 – 15 ml  
      Should have a digital timer  
           

 

 

page 19 of 27

 

    Equipment & Furniture Component  
       
      Body should be made of strong fabricated & corrosion resistant steel
      Control panel – for start/stop switch, dynamic brakes, step less speed regulator with
    1 zero start switch & speed indicator with timer and protective fuses.
      Door interlock  
      Maintenance-free brushless drive motor with exact speed pre-selection and display.
      Speed range 100 to 5000 rpm and above, accuracy 1 rpm.
      System Configuration Accessories, spares and consumables
      Centrifuge complete with Swing and basic rotors , 8 seats , and buckets
      Tube Holders as appropriate , 15 ml Tubes  
      Environmental factors :  
      Shall meet IEC-60601-1-2: 2001(Or Equivalent BIS) General Requirements of
      Safety for Electromagnetic Compatibility.  
      The unit shall be capable of being stored continuously in ambient temperature of 0-
      50 deg C and relative humidity of 15-90%  
      The unit shall be capable of operating continuously in ambient temperature of 10-40
      deg C and relative humidity of 80%  
      Power Supply :  
      Power input to be 220-240VAC, 50Hz .  
      Standards  
      CE;  
      EC Marked  
      US FDA;  
      ISO certification  
       
      Incubator 37° C – Outer case, inner case and front door manufactured with 18/10
      stainless steel, insulated with high grade glass wool.
      Door with silicon gasket and lock key. Front see-through inspection window.
      Heating elements shielded with 18/10 stainless steel.
      Adjustable mesh shelves manufactured with 18/10 stainless steel. Forced air
39 Incubator laboratory 2 ventilation.  
Thermostat  
       
      Thermometer with analogical dial. Safety thermostat.
      Possibility of continuous running : 0-120 min.  
      Capacity not less than 60 liters  
      Power Requirements – 220 ± 10% V. 50 Hz.  
         
      For medicine storage.  
      Upright Type  
      Internal volume not less than 250 lt. / freezer – 40÷50 lt.
40 Refrigerator 3 Temperature controlled within the range of +2°÷ +8° C.
    Number of shelves – 3 or 4.  
       
      With defrosting system  
      Power Requirements: 220 ± 10% V. 50 Hz.  
      Horizontal autoclave, benchtop Type for CSSD, Laboratory , OutPatient/Wards
      Sections  
      Self-contained ,front-loading and fully automatic Unit with integrated Steam
      Generator .  
      The Unit shall be single Manual Door Type  
      Stainless Steel Chamber and Door : Z3CN18/10  
      Capacity per cycle shall be at least 100L for a Cycle time of not more than 25 Mn.(
      Sterilization Phase at 134°C).  
      Sterilization shall be enhanced by the use of Vacuum Cycle Digital temperature
      control and timing , Pressure Control , fully programmable cycle parameters
41 Autoclave   Low Water alarm, Door Safety Locks, Program Cut-Out in case of Power Failure ,
      Automatic Zeroing. Cycle End Alarm  
    2 During Cycle progress , Door is locked into position by both Temperature and
      Pressure Interlocks . Rapid single action door closure shall be of concern
      Electronically-operated Vent valve , Automatic Air Purging , Over Pressure Safety
      Valve and Low Water Cut-Out with Visual and Audible Alarm indication shall be as
      standard  
      The Unit shall be equiped with at least 4 Stainless Steel Trays
      Power Supply: 220VAC.50Hz .  
      Accessories :  
      4 Stainless Steel Trays  

 

 

 

page 20 of 27

 

    Equipment & Furniture Component  
         
      Standards  
      CE;  
      EC Marked  
      US FDA;  
      ISO certification  
      Electronic Controlled Sterilizing Process  
      Outer case and internal chamber – constructed from stainless steel.
      Two or three adjustable shelves.  
      Wall and hinged Door Insulated with High Grade Glass Wool.
42 Hot air sterilizer 3 Door with Lock and Flat Key.  
  Temperature: adjustable from +50ºC to +200 ºC ±10 %.  
       
      Timer Range: 0-250 minutes. ±10 %.  
      Capacity: 50- 60l.  
      Power requirements  
      220 VAC ±10 %,  50Hz.  
  Scale for Adults 1 Capacity-5-150kg.  
43 Graduation-100gr.  
     
      Precision-25 gr.  
      Washable surface.  
      Manual Resuscitator, Autoclavable,with unidirectional Valve
      The supply of Filters shall permit the use in Toxic Atmosphere
      Adapted to First Emergencies the unit shall also the complement to all automatic
      Ventilation  
      Designed for Adults . Shall be supplied with 2 Masks  
      All parts shall be autoclavable up to 134° C.  
      The device shall consist of a self-filling Bag of thin Silicone Rubber which shall
      transmit adequately the User’s Fingertips Action with a high degree of ‘feel’
      indicative of Pressure and Volumes Changes  
      An inflating valve ( Transparent Polysulfone ) shall be fitted between the orifice of
      the Bag’s Delivery End and 2 Rubber Face Masks shall be supplied
      Designed Leak shall counteract Inflating Pressure for all Lung’s Conditions
      A pressure relief valve shall be incorporated to comply with BS 6850.1987
      Balloon : Silicon 1,5L  
    1 Safety Pressure : 30 cm H2O  
44 AMBU emergency case Accessories :  
   
      Balloon , autoclavable  
      Valve, autoclavable  
      Valve Obturator  
      PEP Valve Adaptor  
      2 Masks vinyl, Adult / Children with Straps  
      Tube for Pressure retake  
      Plastic carrying case  
      Standards:  
      CE;  
      EC Marked  
      US FDA;  
      ISO certification  
      Bacterial Lamp for Air Sterilization by means of UV Beams , UVC Category or Far
      UV (200nm< >280nm) with Strong bactericide effectiveness with a maximum
      strength situated at 265nm (253.7nm Mercury Spectra )  
      UVC shall be emitted thru vacuum ionizing Tube ( Weak permeation )
      Useful power shall be at least 45% of nominal power (30W)
      LifeTime shall be at least 7500 Hrs  
45 Bacteriological light 15 Radiation shall be witnessed by means of blue Light emission
Encased in Steel Sheet casing , Epoxy-painted. Ceiling or wall fixation
      Power supply:220V.AC50 Hz.  
      Accessories/Spare Parts:  
      Ionized Tube : 5  
         
46 Infusion fluid holder      

 

 

page 21 of 27

 

    Equipment & Furniture Component  
           
    20 IV Pole , Mobile on Castors    
      Adjustable height    
      2 Hooks    
         
      Manufactured from chrome plated steel tube  
      Upholstered fixed Top covered with washable plastic material , Flame Retardant
      Complete with side Rails and IV Stand  
47 Patient trolley 5 Mounted on 4 casters 100 mm two of which with brakes  
    Knock down construction.    
         
      Length min. –2000 mm.    
      Height – 700 mm.    
      Nursing incubator mobile on castors  
      Temperature control by air temperature or skin temperature  
      Temperature control range 20-37 °C  
      Display of temperature    
      Alarms    
      High and low air temperature    
      High and low skin temperature    
48 Incubator neonate   Fan failure    
  Sensor failure    
      Power failure    
      Secondary thermostat    
      Heater power indicator    
      At least 2 hand ports    
      At least 2 tubing ports    
      Provision for oxygen supplementation  
      Provision for humidity supplementation  
      Internal noise < 60 dB(A)    
      Power supply 220 V 50 Hz    
      Mobile, Stand-type.Max. Light intensity (Ec) up to not less than 110
      000 lx or more Color rendering index (general) Ra 94-96 Color
      rendering index (red) R9> 90 Light field size approx. 13-25 cm. Co lour
49. Surgical light mobile   temperature 4200º K (approx.) Working range without focusing
  intensity approx.70 – 130 cm Power supply 220 V, 50 Hz Light source 3
     
      x halogen approx. 23 V /50 W Durability of halogen bulbs 1300 h or
      more Adjustment of light intensity approx.50 – 100% (0-100%)
      Sterilizable handle    
50. Splint set   Available    
51. Transportation splint   Available    
52. Stretcher   Available    
54. Tracheostomy set   Available    
      Optical System    
      Field of view: 100°÷ 120°    
      Depth of field: 3÷50 mm.    
      Distal end    
      Bending Section    
      Range of tip bending: not less than Up 180°, Down 130°.  
      Insertion tube    
      Outer diameter: not more than 5.0mm  
      Working length: not less than 600mm.  
      Total lenhgth: not more than 900mm.  
55. Fiber bronchoscope   Instrument channel    
  Inner diameter: not less than 2.0mm.  
       
      Light Source    
      Mobile    
      Lamp    
      Type:Quartz halogen.    
      Output: 150W.    
      Cooling fan prolongs lamp life not less than 500 hours.  
      Air/Water functions available    
      Power consumption:not more than 250 W.  
      Power Requirements – 220 ± 10% V. 50 Hz. standard network
      Weight: not more than 6 kg.    
           

 

 

page 22 of 27

 

    Equipment & Furniture Component  
       
      Telescope: 1.2mm , 0º direction of view. Trocar Tube: 1.2mm Trocar
      spike: 1.2mm Accessories: Adapter with Luer lock male and x-mas tree
      Telescope holder Light quid, 3.5m with condenser Trocar Tube for
56. Arthroscope   surgical instruments 2mm Trocar spike 2mm pointed Trocar spike 2mm
  blunt Palpation Probe 2 mm straight -4pcs. Palpation Probe 2mm Hook
     
      Type -4pcs. Knife 2mm side cutting -4pcs. Knife 2mm front cutting –
      4pcs. Rasp 2mm -4pcs. Biopsy Forceps 2mm -4pcs. Grasping Forceps
      2mm -4pcs. Instruments Tray Shaver System
      Optical System Field of View: 80º minimum Direction of View: Lateral,
      Rear-view 15º minimum Depth of Field: 5-60 mm minimum Distal End
      Outer Diameter: max. 12.0mm Instrument Channel Inner Diameter:
57. Duodenoscope   minimum  2.8mm  Insertion  Tube  Outer  Diameter:  max.  11.0mm
      Working Length: minimum 1235mm Bending Section Range of Tip
      Bending minimum: UP 120º, Down 90º Right 110º, Left 90º Biopsy
      Forceps Minimum Visible Distance: 10 mm from Distal End
      Laparoscopy Surgery Set Including:
      Forward-Oblique Telescope 300 deg, diameter 10mm, length 31cm,
      including  
      fiber optic light transmission incorporated. Color code: red.
      Forward-Oblique Telescope 300 deg, diameter 5mm, length 29 cm,
      including  
      fiber optic light transmission incorporated. Color code: red.
      VERESS Pneumoperitoneum Needle with spring loaded stylet LUER-lock,
      length 13 cm.  
      Trocar 11mm with Cannula and multifunctional valve.
      Trocar 6mm with Cannula and multifunctional valve.
      Reducer 11/5mm.  
      Reduction Sleeve, reusable, instrument diameter 3mm, trocar cannula O.D.
      6 mm.  
      Reduction Sleeve, reusable, instrument diameter 10mm, trocar cannula O.D.
      15 mm.  
      Reduction Sleeve, reusable, instrument diameter 5mm,trocar cannula O.D.
      11 mm.color code:green  
      Dissecting and Grasping Forceps, rotating, with connector for unipolar
      coagulation size 5mm, length 36 cm.
      KELLY and Grasping Forceps, rotating, with connector for unipolar
      coagulation size 5mm, length 36 cm.
  Laparoscopy system with   KELLY Forceps insert, long  
58. instrumentation for   MANHES grasping forceps with connector for unipolar
  general surgery   coagulation 5mm, length 36cm  
      METZENBAUM scissors, rotating, with connector for unipolar
      coagulation 5mm, length 36cm,blades curved, double action jaws, length of
      blades 12 mm.  
      Insert METZENBAUM scissors  
      Hook Scissors, rotating, size 5mm, length 36cm,single action jaws
      Coagulating and Dissecting Electrode, L-shaped, with connector pin for
      unipolar coagulation size 5mm, length 36 cm
      Coagulating and Dissecting Electrode, spatula-sheaped, blunt with
      connector pin for unipolar coagulation size 5mm working length
      Insert Micro-dissecting scissors  
      BLAKSSLEY Forceps insert  
      Suction and coagulation cannula, L-shaper
      Take-apart Bipolar Coagulating Forceps
      Insert Forceps  
      Applier for ligating clips (medium large) dismanting, rotating
      Pilling-Titanium-Clips, medium-large, sterile box with 16 cartridges,
      10clipseach  
      Electronic CO2 Endoflator, CO2 bottle empty, power supply, 240 VAC, 50
      Hz,  
      CO2/ N2O gas filter for use with Electronic Endoflator
      High Frequency Surgery Unit, power supply: 200-240 VAC, including: 400
      A Mains Cord Adaptor, unipolar HF- output, 4 mm/5 mm

 

 

page 23 of 27

 

    Equipment & Furniture Component  
       
      Connecting Cord, for connecting the neutral electrodes
      Neutral-Electrode made of conductive silicone, with 1 rubber strap for
      fastening contact face, surface area approx. 180 sq.cm.
      Two-Pedal Footswitch, for control of coagulation.
      Monopolar High Frequency Cord with 5 mm plug for HF-unit, length 300
      cm  
      Suction Bottle, 1.5L.  
      Bacterial filter  
      Cold Light Fountain: approx. 175 W power supply: 220-240 VAC, 50 Hz
      Fiber Optic Light Cable, size: 4.8 mm, length 180 cm
      Endoscopic video system, color system PAL power supply: 240 VAC, 50
      Hz.  
      Cover for use with endoscopic video cameras, size 17 x 242 cm, rolled up.
      Sterile, in packs of 15.  
      Color Monitor, color system PAL, NTSC screen 45 cm power supply: 240
      VAC, 50 Hz including: mains cord  
      Mobile  cart  for  video  system:  4  antistatic  dual  wheels,  2
      equipped with locking brakes, 2 fixes shelves, 1 drawer unit with lock, 1 set
      of non-sliding stands for units, inclusive integrated small cable conduit in
      both vertical boom, 1 camera mount.  Power Box socket board with 12
      plugs,  12  grounding  plugs  Approximate  Dimensions:  cart:  700mm  x
      1280mm x 686mm (wxhxd) Shelf 630mm x 480mm (wxd) caster diameter:
      125 mm  
       
      Electroencephalograph- analyzer. Based on the personal computer. Routine
      EEG studies with computerized analysis and topographical mapping, in-
      depth analysis of EEG/EP activity, analysis of epileptiform activity, long-
      term EEG monitoring , differential diagnosis 24-EEG inputs. Sensitivity:
      0.1 µV to 200 µV. Calibration µV: square (1Hz.) or Sine (5Hz) signal of 5,
59. EEG   10 ,20, 50, 100, 200 and more High frequency filter: >70 Hz at highest
  setting) Low frequency filter: <0,3 Hz at lowest setting) AC Rejection
     
      filters (50/60 Hz.): 1:1000 or better. Automatic suppression of miographical
      artifacts: 20 dB (30 Hz.) or better. Digital signal processing: 22-bit ADC
      Software-controlled Photo and Audio stimulators. Check-up of electrode
      impendence. Power Requirements – 220 ± 10% V. 50 Hz. standard network
      Standard set . Set of standard accessories. EEG-gel 2 kg.
      Gantry  
      Number of slices per rotation-2  
      Gantry aperture: at least 65 cm  
      Gantry inclination angle: at least. 20 degrees  
      Complete scanning time: 1.5, 2.0, 3.0, 5.0 sec app.  
      Min. scanning time: app. 1.5 sec  
      Patient table  
      Coverage region: 70 cm  
      Table height: 40-90 cm  
      Scanning horizontal range: 120 cm  
      Max patient weight: app. 180 kg  
      Acuracy of horizontal movement +/-: 0,25 mm  
      Detector  
60. CT-scanner   Detector type: Solid body  
      Number of detecting elements: at least 708 per row  
      High-contrast resolution: at least 13 par. line/cm at 0% MTF
      Low-contrast resolution: less than 4 mm at 0,3% at 20 mGy
      X-ray subsystem  
      X-ray generator capacity: min 24kWt  
      mA range at least 20-250 mA  
      kVp range at least 80-140 kV  
      Scanning parameters  
      Max duration of spiral scanning: 60 sec  
      Pitch: 0,1:1 to 3,0:1  
      Slice thickness: 1,2,3,5,7,10 mm  
      Min slice thickness: 1mm  
      Summary volume of hard discs: min 80GB  
      Monitor: LCD  

 

 

page 24 of 27

 

    Equipment & Furniture Component  
         
      Monitor size: min 17’’  
      Power requirements: 380V, 3-phase.  
       
      General Description: General purpose portable ultrasound scanner, with
      multiple focus electronic probes. 1. System Specifications: PROBES:
      Electronic 3.5MHz convex Electronic 7.5 MHz linear Scan Modes: B, M,
      B+M, B+B Scan Converter: −Matrix size: min 512x512x8 −Gray-scale
      levels: min 256 Software Features: −Multiple beam focusing −Image
  Portable ultrasound   preprocessing −Image postprocessing −Zoom −Digital calipers −General
61.   analysis: distance, area, circumference −Obstetrics software
scanner  
    packageKeyboard: Alphanumeric keyboard with pointing device (joystick
     
      or trackball). Maximum Display Depth, cm: 22 or more Monitor: High
      resolution monochromatic, minimum size: 7-inch Connections: -standard
      video output 625 lines / 50 Hz -image freezing footswitch Power
      Requirements: -Mains: 220V, 50Hz -Rechargeable battery, minimum
      capacity: 1 hour WEIGHT: maximum 15 kg  
      Type : Galilean Magnification : 4.5X, 7.5X, 11.5X, 18.5X, 29.5X or similar
      Field of View : Ø 49 mm to Ø 8 mm Illumination : Achromatic shadow
62. Surgical microscope for   free spot Light Source : Cold light through fiber-optic light guide Bulb :
eye   Reflector type halogen bulb 15V 150W (24V 250W optional) Illumination
   
      field : 48 mm (Ø) Light intensity : 80000 LUX max. Filter : Various filters
      are available  
      Infant Warmer for Intensive Care Procedures, Includes Micro-processed
      Temperature Control and Orientable Halogen Examination Lamp, with
      Lateral Movement Allowing Free Access for X-Ray.
      Crib of Wide Dimensions 55×65 cm approx. with Washable and Hypo-
      allergic Mattress Tray Apt for X-Rays.  
      Lateral Panels Foldable up to 180º, with Four back IV Ports.
      Trendelenburg and Fowler Movement of ±15º minimum
      Displays:  
      Digital Skin Temperature Thermometer  
      Resolution ≥ 0.1ºC, Range minimum 20 to 42 ºC  
      Temperature Accuracy ±0.3ºC or better  
      Digital Set Point, Resolution 0.1 ºC, Controllable Range minimum 34 ºC to
      38 ºC  
      Heater Power Indicator  
      Skin or Manual Control Mode Indicator  
      Performance  
      -Continuous Circuit Monitoring  
      -Self-calibrating Temperature Circuits  
      -Automatic default Set-point of 36 ºC for Skin Temperature and
      Memory for Control Temperature  
63. Infant reanimation centre   -Memory for Control Set-point Short Term Energy Failure
      -Interchangeable Skin Temperature Probes: ±0.1ºC
      Alarms  
      Skin Temperature ±1ºC  
      Patient Alarm: Heater Remains at max. Power for more than 15
      Consecutive minutes.  
      Skin Temperature Sensor Failure.  
      Electronic Circuit Failure  
      Power Failure  
      Alarm Reset and other  
      Accessories:  
      Skin Temperature Sensor  
      Lateral Rail for Accessories  
      IV Pole  
      X-Ray Cassette Tray  
      Patient Identification Card Holder  
      Monitor Shelf  
      Auxiliary Connection Box  
      Holder for Puls-Oximeter  
      Puls-Oximeter for Infants  
      Lateral Handles for Moving and Instruction Manual

 

 

page 25 of 27

 

    Equipment & Furniture Component  
         
      Phototherapy Lamp  
      Resuscitation System  
      Power requirements  
      220 VAC ±10 % , ≤50W, 50Hz.  
      Twins monitoring capability  
      External ultrasound transducers for measurement of foetal heart rate
      Detection of foetal heart echo by autocorrelation  
      Foetal heart rate tange at least 50–210 /min  
64 Cardiotocograph   Alarms for high and low foetal heart rate  
      External displacement transducer for measurement of uterine activity
      Digital and graphical display of foetal heart rates and unterine activity
      Two channel recorder with event marker and annotation capability
      Power requirement 220 V, 50 Hz  
      Nebulizer with compressor on mobile stand  
      Aerosol particles in range 1 to 5 micron  
65 Nebulizer   Flow rate at least 4 ml/minute  
      Liquid/medication container at least 200ml  
      Power requirement 220 V, 50 Hz  
      High frequency generator  
      kV range at least 22–35 kV with 1 kV increments  
      mA range up to at least 100 mA  
      Time range at least 0.05–86sec  
      mAs range at least 5–500 mAs  
      Automatic exposure control  
      X-ray Tube  
66. Mammograph   rotating anode  
      heat capacity at least 300,000 HU  
      heat dissipation rate at least 60,000 HU/min  
      target/filter Mo/Mo, Mo/Rh  
      dual focus 0.1/0.3 mm  
      Breast compression system with automatic, manual and fine adjustment
      Grid ratio 5:1  
      Radiation shield with lead glass as required  
      Single plane angiography system  
      4-way table top motion  
      Collision prevention system  
      X-ray Generator  
      high frequency  
      100 kW @ 100 kVp  
      radiographic mA up to 1000 mA  
      radiographic kVp 60–150 kVp  
      radiographic timer up to 1.5 sec  
      fluoroscopic mA up to 10 mA  
      fluoroscopic kVp 60–110 kVp  
      pulsed fluoroscopy  
      cine range up to at least 30 f/sec  
  Coronarography/Angiogr   X-ray tube  
67.   80 kW  
aphy system    
    heat storage capacity at least 1,500 kHU  
       
      heat dissipation rate at least 300 kHU/min  
      dual focus, focal spot size 0.6/1.2 mm or less  
      grid pulsed and cooled  
      Digital dynamic flat detector  
      23 cm with zoom fields  
      Storage capacity at least 50,000 images at 1,024 × 1,024
      At least two 17 inch monitors  
      Digital subtraction angiography  
      Cardiac analysis software package  
      Other software packages to suit user applications  
      DICOM 3 compatibility  
      Radiation shield as required  
      Contrast injector  

 

 

page 26 of 27

 

    Equipment & Furniture Component  
         
      Monitoring system  
      Accessories as specified by user  
      Room preparation as required by user  
      Convertible into operating table for emergency Caesarian section
      Adjustable backrest  
      Removable headboard  
      Leg supports  
68. Delivery Couch   Removable foot section  
      Height adjustable over at least 60-90 cm  
      With castors at least 12 cm dia, at least two diagonally lockable
      Corner bumpers  
      Removable drain pan  
      Heat control  
      manual or automatic  
      range includes 34–37 °C  
      heater indicator  
      Temperature display  
      set point temperature  
69 Infant warmer   skin temperature  
      range includes 30–39 °C  
      Alarms  
      audiovisual  
      sensor disconnect  
      power failure  
      Power requirement 220 V, 50 Hz  
      Capacity at least 50 kg/cycle  
      Cylinder volume at least 500 litre  
      Stainless steel construction of parts exposed to corrosion
70. Washing machine   Electric heating  
      Adjustable extraction speed including 300–500 rpm  
      Multiple liquid supply and detergent connections  
      Programmable  

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

page 27 of 27