 LIST OF MEDICAL EQUIPMENT 
    
ItemNameQuantityTechnical Specifications and Standards
No    
    
   Processing machine for X-ray films from 13×18 cm. 18×24 cm.24×30
   cm.30x40cm.35.6×35.6 cm. 35x43cm. 
   Capacity at least 60 films per hour 
 X-Ray Film processor Standard Equipment: Processor + set of 3 replenish bottles 
1 Accessories: Water supply connection with shut-off valve , Filter , Light-tight
Table top 
 1cover , Processor stand 
   
   Automatic stand-by mode. 
   Automatic film detection 
   Power Requirements : 
   220 VAC ±10 % 50 Hz . 
     
   X-ray Film Illuminator (Viewer) . 2 Fields . Wall-Mount 
   Diffuse, uniform, flicker free illumination 
   Transparent Spring-loaded film Retainers shall grip lightly and firmly without
   obscuring top edge details 
   Screen shall be recessed into the cabinet to help keep the interior dust free and
   eliminate side light spill 
   The film retainers shall always operate effectively maintaining an even pressure
2X-ray Film Viewer across the full width of the illuminator. 
 Direct Starting , Luminous Source : Daylight 
    
   Command and Control on Front Side. 
   Bipolar Switch with Pilot lamp 
   Wall Fixations shall be provided for . Cord , Local Plug 
  8Power Supply : 220V/50Hz 
   CE; 
   EC Marked 
   US FDA; 
   ISO certification 
   Universal, Remote controlled Universal Unit 
   Processor controlled, Mixed Cassette’s Radiography and Digital Fluoroscopy
   screening System. 
   Overhead Tube on the freely moving Tube Arm without floor mounted column.
   The System to be suitable for Standard Skeletal and Radiographic Examinations,
   Including Lateral Exposures and Oblique Beam Projections
   Fully Automatic Under Table Spot Film Device for Cassette Radiography with
  1Extensive Range of Cassettes, Grids and Attachments for Image Intensifier.
   Free Cassette Exposures on Table, Floor, Wheelchair or Gurney.
   Table. 
   Tilt: Motor Driven, +90º to -15º, automatic stop in horizontal position
   Height: approximately 85 cm. 
   Tabletop Outside Dimensions: length approximately 200cm,Width approximately
   80cm 
3X-Ray Unit Universal Radiolucent: not less than 190cm x 55cm. 
   Longitudinal Travel: Motor Driven, at least 160 cm 
   Transverse Travel: at least 20 cm 
   Patient Weight: at least 200 kg. 
   Tube Assembly. 
   Max. exposure voltage : 150 kV 
   Anode heat dissipation rate at least 30 kW for small focus and 60 kW for large
   focus 
   Anode heat storage capacity at least 500 kHU 
   Dual focus 0.6 mm and 1 mm 
   Complete filtration W ≥ 2.5 mm Al 
   Focus-Film Distance (SID) Fixed Variable : 115cm / 150cm. 115cm and 150cm
   must be set by Motor Driven with Adjustment Speed 
   Oblique Projections: max.-40º-+40º ( SID 115cm.) and -35º-+35º (SID 150cm),
   Automatic Parallax Compensation Between Cassette and Image Intensifier Input
   Screen in Central Ray. 

 

page 1 of 27

 

Equipment & Furniture Component

Tube Assembly Swivel: Manually in the Range +90º to -90º with Stops max. every 10º and -90º to 180º with stops max. every 30º.

Spot Film Device.

Front loading, automatically drawing in and out, centering and format sensing for

Cassettes, of the Formats 18x24cm (8″x10″) to 35 x 43cm (14″x17″), Standardized

 

According to IEC, ANSI and DIN

Automatic Format Collimation: must be Selected Separately According to Format

Height/Width for Cassette Spot filming; Automatic Formatting for Bucky

Exposures, Object-related Collimation.

 

Film Segmentation: Minimum 4 on 1., in Multiple Segmentation According to

Cassettes Program

Inward Movement Time: Park to Exposure position not more than 1sec.

Time Interval: Fluoroscopy/Radiography ≤1.0 sec. without grid movement.

 

Scattered Radiation Grid: Stationary 17:1, 70 Lines/cm, Fo=125cm.,

Excursion: minimum 105cm, Remote-controlled.

Compression Device.

 

Radio transparent, Remotely Controlled, Detachable and Replaceable Cone minimum 3 Shapes.

 

Compression Force range: from 5 to 155N, Movement Blockage starting from 50N Force Indication: Digital at the System Remote Control Console.

Projection Angle: -30º-+30º.

 

Image Intensifier and TV System

3 Image Fields 13 to 23 cm

Visual resolution: Mean Value 5.2; 5.8; 6.4 Lp/mm

 

Television System Standard: Line Frequency 50Hz; 625 Lines with 50Hz. TV Technology : CCD sensor, noise suppression, LIH function Automatic Dose Stabilization (ADR)

 

Monitor: 44cm or more, Frame Rate at least 50 Hz.

 

X-Ray Generator

 

High Frequency, Multipulse Up to 150 kV, ≥50kW.

 

Automatic System: 1-Point Technique with Continuously Falling Load; 2 -,3- Point Technique with Constant Load; 3-Point Technique with AEC

 

Organ Programs: minimum 20 Organ Programs must be generated and stored. Number of Workstations: not less than 5 of with 1 Fluoroscopic Workstation Fluoroscopy working range: minimum 40kV/ 0.5mA up to maximum 110kV≤10mA.

 

Fluoroscopic Mode Selectable: Fluoroscopy Under Manual Control / Automatic Control

 

Fluoroscopy Under Manual Control: manual select kV but mA must be calculated from Anti-isowatt, curve

 

Fluoroscopy Under Automatic Control: Minimum 2 required Characteristic Curve must be selected for Automatic Dose Rate Control : us Minimal Dose/ Contrast Litho ( or other according customer ordering)

 

Tube Connection: minimum 2 Double-Focus Tube Assembly. 3 field Automatic Exposure Control

 

Primary Collimator: Inherent Filtration 1.0mm Al; additional Filters 0.1mm Cu, 0.2mm Cu, 0.3mm Cu.

Ambient Conditions

 

Operation: +10Cº to 35Cº; 20% to 75% relative humidity, non-condensing; 70kPa to 106kPa

 

Storage and Transport: -20Cº to+70Cº; 10% to 95% relative humidity, non-condensing; 70kPa to 106kPa

Power Connection

 

Nominal Voltage: 3/N/PE, 380V±10%

Nominal Frequency: 50Hz.

Completeness:

 

Lead glass, radiation protection window 80x100cm. ±10 % , 2,1 Pb equiv X-ray protection:

 

1x Coat apron, 0.35 Pb, size medium, 110cm. X-ray protection coat apron for general examinations.

 

1x Gonad protection apron 40x37cm, 0.5 Pb. Apron for adults for protection of ovarian and gonad.

 

1x Gonad protection capsule for boys, 1.0 Pb.

  • Set of cassettes
  • Cassette without patient ID window
  • Fitted with green screen

2x Cassette with green screen, 18 cm x 24 cm

 

 

page 2 of 27

 

    Equipment & Furniture Component  
         
      1x Cassette with green screen, 18 cam 43 cm  
      2x Cassette with green screen, 24 cm x 30 cm  
      2x Cassette with green screen, 35 cm x 43 cm  
      1x Cassette with green screen, 35 cm x 35 cm  
      Exposure range at least 0.5–200 mAs  
      Shortest exposure time: not more than 10ms.  
      Max. exposure time: not less than 5 sec.  
      Tube head must be rotatable in all planes.  
      Adjustable collimator and light marker Easy to use and to move. Non automotive 
      Two point technique – kV and mAs.  
      High- frequency X-ray generator with multipulse voltage provider. 
4 X-Ray Mobile 1 The touch-screen keys and digital display of kV and mAs values. 
      Output: not less than 2.5kW. 120 kV, 40mA.  
      Focal spot not more than 1 mm.  
      Collimator: rotatable ± 90°  
      Max. horizontal extension at least 800mm  
      Power Requirements  
      220 V ±10%, 50Hz  
      Wide Dynamic Range Digital Front-end  
      The 12-bit A/D converter  
      Multi-beam processing.  
      Image Display Modes  
      B: gray-scale imaging, M,D: Spectral Doppler (PW,HPRF PW), 
      Flow: Color Doppler and Power Flow imaging, Dual B, B and M, B and D, B(Flow), 
      Dual B (Flow), M(Flow), B(Flow) and M(Flow), B(Flow) and D, 
      Triplex mode: B, Flow and PW Doppler simultaneous real-time display, 
      Dual Dynamic Display (DDD): B and B (Flow) simultaneous real-time display, TDI 
      (Tissue Doppler Imaging)-Available on electronic scanning probes. 
      Request function: In multi-mode display, it is possible to select one image for full 
      screen display.  
      DMS (Integrated Data Management System)  
    1 Measurement and Analysis.  
      General measurements  
      On B-mode image  
      On M-mode image  
      On spectral Doppler  
      On B/D-mode image  
      On B(Fiow) mode image  
      Obstétrical measurements & calculations  
5 Ultrasound (3 probes)   Gynecological measurements & calculations  
    Peripheral vessels analysis  
        
      Urological measurements & calculations  
      Abdominal measurements  
      Report Functions  
      Data Communication Function  
      Preset Function  
      Characters and graphic displays  
      Probe connectors  
      Standard  
      Active connectors: 2 connectors  
      Dummy connector for rest: 1 connector  
      Viewing Monitor  
      15-inch diagonal color Liquid Crystal Display with brightness and contrast controls 
      Acoustic Power  
      0 to 100%, changeable  
      Environmental Requirements  
      Temperature: +10 to +40 degrees C  
      Relative Humidity: 30 to 75%  
      Power Requirements: 220V / 50 Hz.  
      Probes  
      General abdomen OB/GYN 2.5-6.0 MHz.  
      Small parts, PV (steered linear) 5.0-10.0 MHz.  
      Adult heart (harmonic echo) 2.1-3.8 MHz.  
6 Gastroscope (With halogen   Optical System :  
 light source)   Field of view: at least 100°°  
       
      Depth of field: at least 4 to 100 mm  

 

 

page 3 of 27

 

Equipment & Furniture Component

 

   Distal end : Outer diameter: not more than 10mm.
   Bending Section
   Range of tip bending: not less than Up 210°, Down 90°, Right 100°, Left 100°
   Insertion tube : Outer diameter: not more than 13.3mm
   Working length: not less than 1025mm.
   Length markings
   Instrument channel :
   Inner diameter: not less than 2.8mm.
   Light source
   Halogen technology shall provide a high light intensity of at least 150W.
   Variable light intensity.
   Fast and easy bulb change during operation : In case of a bulb failure immediate
  1switch to the second bulb with the reverser shall be possible.Adjustment of
   brightness and power with buttons and switches on the front panel.
   Shall be connectable with conventional telescopes, fiberscope.
   Shall be adaptable to all endoscopic procedures
   Optionally , an automatic built-in swivel mechanism will bring the spare halogen
   bulb into working position.
   Technical Data : Halogen , 150 watts , Halogen, 2 pieces
   Manual switch , 2 x 150 W , 24 V
   Power Supply : 220 VAC 50 Hz .
   Standards
   CE;
   EC Marked
   US FDA;
   ISO certification
    
   Optical System :
   Field of view: at least 120°
   Depth of field: 5 –÷100 mm.
   Distal end : Outer diameter: not more than 13mm.
   Bending Section
  1Range of tip bending: not less than Up 180°, Down 180°, Right 160°, Left 160°
   Insertion tube : Outer diameter: not more than 13.3mm
   Working length: not less than 1680mm.
   Instrument channel
   Inner diameter: not less than 3.2mm.
   Light Source
   Halogen technology shall provide a high light intensity of at least 150W .
   Variable light intensity
 Colonoscope Fast and easy bulb change during operation : In case of a bulb failure immediate
7(With halogen light switch to the second bulb with the reverser shall be possible
  
 source) Adjustment of brightness and power with buttons and switches on the front panel.
   Shall be connectable with conventional telescopes, fiberscope,
   Shall be adaptable to all endoscopic procedures
   Optionally , an automatic built-in swivel mechanism will bring the spare halogen
   bulb into working position.
   Technical Data : Halogen , 150 watts , Halogen, 2 pieces
   Manual switch , 2 x 150 W ,24 V
   Power Supply : 220 VAC 50Hz .
   Standards
   CE;
   EC Marked
   US FDA;
   ISO certification
    

 

 

page 4 of 27

 

  Equipment & Furniture Component 
     
   Recording ECG Leads: 12 standard Leads 
   Recording Channels: 3/1 user selectable 
   LCD display of ECG 
   Lead switching: manual and automatic. 
   Sensitivity, mm/mV: 5, 10, 20. 
   Calibration signal: automatic and manual. 
   Diagnostic frequency range 0.67–150 Hz or better 
   Filters for mains frequency, low frequency, muscle artefact, high frequency
   Recorder: 
   Recording method: thermal paper 
   Recording speed, mm/sec: 25/ 50 user selectable 
   Channels acquired simultaneously:1/ 3 user selectable 
   Channels printed simultaneously:1/ 3 user selectable 
   Other features: 
   Portable 
  3Mains and internal rechargeable battery operation 
 Electrocardiograph 3Battery operating time, minimum: 90 min. 
8  
 Power requirements: 220 V, 50 Hz 
channels  
    
   Accessories: 
   1. Patient cable 
   2. 6 chest limb electrodes 
   3. 4 limb electrodes 
   4. 4 strap electrodes 
   5. 
   6. 1 bottle ECG Gel 
   7. 2 rolls of paper or Z-Fold 
   8. Carry bag 
   Standards 
   CE; 
   EC Marked 
   US FDA; 
   ISO certification 
    
  3Surgical Lighthead with five Projectors , Ceiling Suspended
   Light intensity (Ec) at least150 000 lx 
   Color rendering index (general) Ra 96 
   Color rendering index (red) R9 89 
   Light field size 15 – 30 cm. 
   Colour temperature 4200º Kelvin 
   Working range without focusing approximately 70 – 140 cm
   Power supply 220 VAC, 50 Hz 
   Adjustable light intensity 
   Power Supply , 220V/50 Hz. 
 Operating theatre light (5 Power Stabilizer shall be supplied 
9 This equipment shall be supplied with all necessary fixation gears to the Building
spotlights) 
  structure 
    
   Replacement Bulbs (5) and Sterilizable Handles (5) shall be supplied
   Standards 
   CE; 
   EC Marked 
   US FDA; 
   ISO certification 
   Accessories/Spares : 
   5 Bulbs per unit 
   2 Handles per unit 
   Operating table designed for carrying out treatments, dressing interventions and
 Operating Table operations of general surgery and traumatology 
10(Trauma/Multifunction, All exposed metallic parts shall be made from stainless, acid proof steel.
 The table base shall be mobile and shall have central brakes
 electric/hydraulic driven 
  Back rest and leg rest inclination angle, kidney elevator, Trendelenburg and
   reverse-Trendelenburg positions and height adjustment of the table top shall be

 

 

page 5 of 27

 

Equipment & Furniture Component

  activated by electro – hydraulic system
  The table top shall be translucent for x-rays with 5 separate Sections
  Table top length approximately 2000 mm
  Table top width approximately 600mm
  Minimal height of the table 750mm
  Maximal height of the table 1100mm indicatively
  Trendelenburg at least 25
  Reverse Trendelenburg at least 25º
  Lateral tilt at least 18º
  Back rest inclination angle at least +55º to -25º
  Head rest inclination angle at least 45º up
  Head rest inclination angle at least 20º down
  Power requirements
  220 VAC ±10 % ,  50Hz
 1Accessories:
  Anaesthetic Screen with clamp with telescopic tubes : 1
  Body restraint Strap with clamp : 1
  Padded Shoulder supports : 2
  Padded Leg support with swivel type clamp : 2
  Padded arm Rests 450 -500 mm long with two arm Clamps : 2
  Padded Lateral support with universal attachment Clamp : 2
  Padded rubber mattresses with Anti Microbial agent incorporated into all components
  that assists in Prohibiting growth of bacteria & fungi and easy to clean and maintain
  and of at least of 1” thickness : 2 X5 Sections
  Head rest , I.V drip Stands attachable to the table
  Environmental factors:
  The unit shall be capable of being stored continuously in ambient temperature of 0 –
  50deg C and relative humidity of 15-90%
  The unit shall be capable of operating continuously in ambient temperature of 10 -40
  deg C and relative humidity of 15-90%
  Accessory for traumatology are optional and shall be listed separately
  Standards
  CE;
  EC Marked
  US FDA;
  ISO certification
  Operating table for carrying out treatments, dressing interventions and operations
  of general surgery (3-4 surgeries daily).
  All exposed metallic parts shall be made from stainless, acid proof steel.
  The table base shall be mobile and shall have central brakes
  Back rest and leg rest inclination angle, Trendelenburg and reverse-
  Trendelenburg positions and height adjustment of the table top shall be activated
  by electro – hydraulic system
  The table top shall be translucent for x-rays with 5 separate Sections
  Table top length approximately 2000 mmTable top width approximately 600mm
  Minimal height of the table 750mm
  Maximal height of the table 1100mm indicatively
  Trendelenburg at least 25º
  Reverse Trendelenburg at least 25º
  Lateral tilt at least 18º
 Operating TableBack rest inclination angle at least +55º to -25º
11(Multifunction,Head rest inclination angle at least 45º up
Head rest inclination angle at least 20º down
 electric/hydraulic driven)
  

 

Power requirements

 

220 VAC ±10 % , 50Hz Accessories:

 

Anaesthetic Screen with clamp with telescopic tubes : 1 Body restraint Strap with clamp : 1

  • Padded Shoulder supports : 2

Padded Leg support with swivel type clamp : 2

 

Padded arm Rests 450 -500 mm long with two arm Clamps : 2 Padded Lateral support with universal attachment Clamp : 2

 

Padded rubber mattresses with Anti Microbial agent incorporated into all components that assists in Prohibiting growth of bacteria & fungi and easy to clean and maintain and of at least of 1” thickness : 2 X5 Sections

 

Head rest , I.V drip Stands attachable to the table Environmental factors:

 

 

page 6 of 27

 

Equipment & Furniture Component

 

   The unit shall be capable of being stored continuously in ambient temperature of 0 –
   50deg C and relative humidity of 15-90%
   The unit shall be capable of operating continuously in ambient temperature of 10 -40
   deg C and relative humidity of 15-90%
   Standards
   CE;
   EC Marked
   US FDA;
   ISO certification
    
   Anesthesia Machine used for delivering anesthesia Agents to Patient during Surgery
   Procedures. The complete set-up shall also ventilate the Patient
   Patient monitoring system  ECG, EtCO2, Pulse Oximeter and airway pressure, NIBP,
   rectal/&skin Temperature.
   The full System shall be completed with :
   Anaesthesia gas delivery system, O2/N2O/Compressed Air .
   Circle absorber system.
   Precision vaporiser for halothane
   Anaesthesia ventilator
   Patient Circuit
   Anesthesia gas delivery system:
  3Shall have provision for delivery of oxygen, nitrous oxide and medical air with
   pressure gauges.
   Oxygen and Nitrous oxide should be linked either mechanically or pneumatically to
   ensure a minimum of 25% oxygen delivery at all times to avoid delivery of hypoxic
   mixture.
   Shall have audio-visual oxygen Failure Warning System with Nitrous oxide cut off.
   Shall have back bar ISO pin type to attach Vaporizer easily.
   Flowmeter oxygen : 0÷2; 2÷10 litres/min, minimum range
   Flowmeter N2O : 0÷2; 2÷10 litres/min, minimum range with safety mechanism
   (hypoxic link) or similar
   Flowmeter , Medical Air 0÷10 litres/min.
   Shall be supplied with necessary Attachments for use of the breathing Circuits
   (Ruben , Bains, Jackson-Rees or Magill )
   Construction :
   Shall have top Shelf to keep monitors and a Tabletop with Drawers to keep
   anaesthetic drugs, equipments etc.
 Anesthesia trolley Castor wheels should be durable , moisture resistant and antistatic
  The Anaesthesia machine frame shall be made of rust proof material/Stainless steel
12(anesthesia machine with 
 or sheet Steel protected with Epoxy-Paint
 ventilator and monitor) 
  Standard Circle Absorber System
   
   Shall have adjustable pressure limiting Valve, breathing circuit pressure measuring
   device.
   Shall have a Bag/Ventilator selecting Valve integrated onto the Absorber.
   Facility to attach oxygen sensor
   Shall have CO2 absorbent Chamber Canister (Soda Lime )
   Precision Vaporizers (Temperature, Pressure and Flow compensated) for Halothane
   Should be easy to mount and dismount from the back bar.
   Vaporisers shall have ISO pin type (Selectatec) Mounting and Vaporiser interlocking
   facility
   Shall have a standard filling port with keyed filling device
   Should be designed for transport with liquid in Vaporiser Chamber with protection
   against tipping and shaking
   Maintenance free Vaporiser
   Ventilator
   Shall be a bag in bottle anaesthesia Ventilator with standing (ascending/Piston)
   Bellows
   Shall be supplied with adult and paediatric Bellows
   Shall be able to set Tidal Volume, Respiratory rate and I:E ratio
   Ventilator should have audible alarms for ventilator failure, low oxygen supply
   Pressure, inadequate volume delivery, disconnection alarm, and power supply failure.
   Ventilation mode: CMV-SV – manual , Minute volume: 2 – 30 l/Mn
   Tidal volume: approx.50 – 1400ml , Breathing frequency: approx.6-60 times/min

 

 

page 7 of 27

 

Equipment & Furniture Component

Flow rate range: 5 – 60 l/min

High pressure relief valve

Power supply 220VAC, 50Hz

 

Accessories :

 

Accessories required for Ventilator :

Resuscitators , Adult / Child : 1 Each

Set of face masks , 3 Sizes : 2 Each

Corrugated hoses and tubing set for Pipeline O2 : 2 Sets

 

Corrugated hoses and tubing set for Pipeline Air : 2 Sets

 

Accessories required for Anaesthesia Machine :

Hose assembly for piped Oxygen supply : 2 Sets

 

Hose assembly for piped Nitrous Oxide supply : 2 Sets

Hose assembly for piped Air supply : 2 Sets

 

Regulator and manometer for the pipe pressure hose ( 60 psig) : 1 Set

 

Connection from ventilator to anaesthetic apparatus , autoclavable/reusable : 1 Set

Test bag, 1 litre : 2

Peak Flow Meter : 1

Power cord with grounding wire : 1

 

Dust cover : 1

Bellows assembly, adult / Children : 2 Each

Patient Circuit , Magill Type , Complete , Adult / Children . . Shall be Reusable : 2

Each

Spirometer (Peak Flow Meter for Anaesthesia procedures)

Consumables for anaesthetic unit for 2 years operation:

Battery cell (as appropriate)

Soda lime for circle absorber (5Kg/pack) : 10 Packs for 1 unit

 

Standards

 

CE;

 

EC Marked

US FDA;

ISO certification

Syringe Driver for Parenteral Infusions , in Wards, OP, ICU’s and during Transportation/Transfer

Single Channel. Shall accept Up to 60cc Syringes , Non-Captive

 

Self checking of Electronics with audible call back on locked-out Controls Tamper and fluid resistant Design . ABS Casing (Impact-Resistant )

 

Shall be fitted with IV Rod, Adjustable Height with 4 Hooks.Mobile on 5 Star Stand with Castors

 

Internal rechargeable battery

 

Rate: 0.1 to 99.9 ml ( in 0.1 ml steps) indicatively Total volume infused: 0.1 to 999.9 ml

Precision : +/-3%.Stipulated Syringes

Occlusion release shall give minimum Bolus effect

 

Syringes : 10 Types at least of LuerLock Type 10/60mL.Syringes,+ Specific Choice

Pause infusion facility

Alarms/Safety

 

13 Syringe pump Infusion nearly complete Infusion complete Occlusion

 

System malfunction Syringe unlocked Plunger disengaged Low battery

 

Protection against Leakage Current:Type CF.Equipment

Protection against Electric Shocks:Class II.Equipment

Safety features: Error coding (event log)

Rate change through Audible Stop Switch

Audible Call back on locked-Out-Controls

  • Automatic detection of Syringe Volume

 

Minimum bolus on Occlusion Release at any selected Rate Communication : RS232C Interface

 

Electrical:

Electrical: Class II (double insulated)

 

 

page 8 of 27

 

Equipment & Furniture Component

 

 

   Bat.Autonomy : 8 Hrs.Minimum
   Power supply: 220 VAC.50.60Hz
   External transformer . Stabilizer
   Environmental factors
   Shall meet IEC-60601-1-2: 2001(Or Equivalent BIS)
   General Requirements of Safety for Electromagnetic Compatibility or should
   comply with 89/366/EEC; EMCdirective.
   The unit shall be capable of being stored continuously in ambient
   temperature of 0 -50deg C and relative humidity of 15-90%
   The unit shall be capable of operating continuously in ambient temperature of
   10 -40deg C and relative humidity of 15-90%
   Accessories
   Coupler between 2 units
   Mobile floor stand
   IV stand clamp
   Rail clamp
   Standards :
   IEC601.1/IP33
   Safety: IEC 601 / 1. CF.
   Compliance with EN 60601.1 and PrEN 60601-1-24.
   CE 0459 marking in compliance with EEC 93/42 Medical Products directive.
   Comply with  EN60601-1-2, EN60601-2-24
   EN ISO 9001:2000, EN ISO 13485:2000; EN 46001:1996
    
   HF electrosurgical unit shall be used to execute monopolar and bipolar surgery in
   many fields of application where high precision and reliability are essential
  3Outputs of cut, coagulate and blend
   Maximum output 300 W for monopolar cut
   Activation :
   Double pedal switch which may be used for the monopolar and bipolar functions .
   Hand-switch handle
   Bipolar electrode with pedal switch or with automatic Start/Stop system ( for
   coagulation only)
   Control
   The Unit shall stop automatically in case of internal error which shall be identified on
   Display and with audible alarm
   Memorization : User shall be able to use at least 4 working programs
   Safety :
   Neutral plate safety circuit shall control connections and contacts of Neutral Plate
   with Tissues:
   Defective Contact shall be notified with visual Alarm and immediate reducing of
   power
   Output circuit : floating – protected against defibrillator interferences . Shall have HF
 Electrosurgical leakages less than 150mA through each electrode
14 Power Supply : 220VAC, 50Hz
Unit (Monopolar-bipolar) 
  Cooling: convection without fan
   
   Accessories :
   Single-use two button Handle , Autoclavable
   Handle shall be provided with a 3 Pins socket that shall fit majority of bipolar
   electrosurgical units
   Operative Foot-switch (usable as alternative to handle)
   Reusable neutral Plate
   Kit of 10 short autoclavable Electrodes
   Electrode wire – straight
   Electrode wire – angled 45°
   Electrode bend – 0 4 – straight
   Electrode bend – 0 8 – straight
   Electrode hook – angled 45°
   Electrode wire 1 mm.- angled 45°
   Electrode straight slip-knot – straight
   Electrode angled slip-knot – straight
   Electrode ball point – 0 3 mm. – angled 45°
   Electrode ball point – 0 3 mm. – straight
   Bipolar Forceps : 6
   Forceps shall be provided with a standard European connection
   Straight Forceps – 18 cm. (7″)

 

 

page 9 of 27

 

Equipment & Furniture Component

 

   Curved Forceps – 18 cm. (7″) 
   Curved Forceps – 20 cm. (7 3/4″) 
   Bayonet Forceps – 18 cm. (7″) 
   Bayonet Forceps – 20 cm. (7 3/4″) 
   Straight Forceps – 20 cm. (7 3/4″) 
   Cable, Bipolar 
   Adaptor, Bipolar Cable 
   User Manual 
   Standards 
   CE;  
   EC Marked 
   US FDA; 
   ISO certification 
    
   Suction Unit for Major Surgery Procedures . Mains-powered , mobile on 4 antistatic
   Castors, ABS Casing and 2 graduated Canisters of 2,000ml each made of
   Polycarbonate autoclavable at 121°C and disposable suction bags 
   Shall require no maintenance nor lubrication 
   Oil-free pump, maximum suction of at least 500 mm Hg 
   Free flow rate at least 25 l/min 
   Main Switch with Pilot Lamp . Fuses 
   Pedal Action 
   Shall be equipped with a protective thermal cut-out relay. 
   Shall be equipped with motor-protection cap that totally prevents aspirated liquids or
   secretions from reaching and damaging the vacuum pump 
   Suction command with continuous adjustment , Vacuometer 
   2X2,000ml Canisters with airproof screwing-cap with independent Overflow devices
   . Fast Connectors and silicone Tubing 
   Power Supply : 220VAC.50Hz. 
   Ventilation Fan for overheating 
15Suction Pump2Sound level: Shall be not more than 55 dBA 
   Accessories 
   Silicone Tubing , sterilizable 
   Transparent Cannula Holder, Sterilizable 
   Anti-Bacterial Filters ( 4 ) 
   Set of 4 canulaes with Holder : Yankhauer , Soft Universal Yankhauer
   Diameter : 8.0/6.0mm with anti-sticking Lumen and High Suction Lumen
   Universal Soft Canulaes diameter : 6.0/4.0mm 
   FrazierCanulaes (Fergusson) diameter :1,5/2.0/3.0/4.0mm 
   Jackson Canulaes : 35×3 /45X3/25X4/35X4/45X4/55X4/60X4cm 
   Standards 
   CE;  
   EC Marked 
   US FDA; 
   ISO certification 
   1Operating Scissors, 170 mm, straight, blunt-blunt points1
   2Operating Scissors, 140 mm, straight, sharp-blunt points1
   3Operating Scissors, 140 mm, curved, sharp-blunt points2
   4Dissecting Scissors, curved, for deep surgery, 230 mm2
   5Operating Scissors, 140 mm, curved, blunt-blunt points4
   6Operating Scissors, curved, blunt-blunt points, 170 mm2
   7Operating Scissors, 140 mm, curved, hardened, blunt-blunt1
   points
     
16Operating instrument set 8Sharp Tips Scissors, curved, 160 mm1
  29Dressing Forceps, standard pattern ПА 150х2,56
   10Dressing Forceps, standard pattern ПА 200х2,54
   11Surgical Forceps, 150×2,58
   12Surgical Forceps, 1 x 2, 200×2.5 mm4
   13Abdominal Depressor, 285 mm, blade 70 mm wide1
   14Tissue Forceps, 265 mm, straight5
   15Tissue Forceps, 256 mm, curved5
      

 

 

page 10 of 27

 

 HSMP ArmeniaEquipment & Furniture Component  HPIU: HH
      
   16Hemostatic Forceps, curved, atraumatic (160 mm)6
   17Hemostatic Forceps, curved, 1×2 teeth, serrated,size 2,6
   (160mm)   
        
   18Hemostatic Forceps, curved, serrated,size 2 , (196 mm)6
   19Hemostatic Forceps, curved, serrated,(240 mm) 6
   22Hemostatic Forceps, straight, 1×2 teeth, serrated,size 1 ,3
   (150 mm)   
        
   23Hemostatic Forceps, curved, 1×2 teeth, serrated,size 1,6
   (150mm)   
        
   24Hemostatic Forceps, curved, 1×2 teeth, serrated, size 1,6
   (200mm)   
        
   25Hemostatic Forceps for deep surgery, child size, size 2,6
   (225mm)   
        
   26Rake Retractor, 2 sharp prongs,(200 mm) 2
   27Rake Retractor, 3 sharp prongs,(200 mm) 2
   28Rake Retractor, 200 mm, 4 sharp prongs, style 2 2
   29Rake Retractor, 2 blunt prongs, 200 mm 3
   30Rake Retractor, 220 mm, 4 blunt prongs, style 2 4
   31Rake Retractor, 6 sharp prongs,(200 mm) 2
   32Abdominal Retractor   2
   33Needle Holder,( 200 mm)   2
   34Needle Holder, ( 250 mm)   2
   35Vascular Needle Holder , 160 mm  2
   36Vascular Needle Holder, 200 mm   2
   37Intestinal Forceps, non-crushing, curved, child size, (192mm)2
   38Intestinal Forceps, non-crushing, curved, adult size, (235mm)2
   39Tongue Spatula, straight, 20х16mm,(180 mm) 2
   40MОSQUITO Hemostatic Forceps, serrated, curved on edge,6
   61/10″ (154 mm)   
   41Intestinal Forceps, non-crushing, straignt, adult size,2
   (240mm)   
        
   42Abdominal Retractor (60 mm)   2
   43Abdominal Retractor (100 mm)   2
   44Liver Speculum (100 mm)   2
       
    
   Suction Unit for Emergency procedures , fitted with single 1000ml capacity
   Polycarbonate Canister , mobile on 4 Castors and carrying handle 
   The Unit shall enable operation from either mains supply, or independently from an
   internal 12V rechargeable battery    
   The unit shall recharge when plugged into the mains  
   Anindicator on the casing shall indicate the status of the battery . A Vacuometer
   shall indicate the suction value    
   Maintenance free diaphragm pump    
   Overflow safety device    
   The Unit shall be capable to generate 500mm Hg at least 
   Free flow rate at least 25 litres per min    
   Shall be fitted with easily replaceable anti-bacterial filter with 5 spare
  1Shall provide over 30 minutes of full vacuum from a fully charged battery
17Suction Pump Fitted with rechargeable battery    
   Operating mains voltage: 220 V – 50 Hz    
   Recharging is by trickle charge and shall take eight hours maximum to restore full
   charge    
   Accessories :    
   Hand piece with regulator and appropriate probes shall be supplied as well as an
   additional Canister    
   Standards    
   CE;     
   EC Marked    
   US FDA;    
   ISO certification    
    
   The following specifications define a portable, lightweight ; battery or accumulator
18Defibrillator operated automated external to treat Patients requiring basic Life Support
 Weight of complete unit, excluding batteries or accumulator , shall permit easy
   
  2Transport . The overall Dimensions shall enable easy storage and portability

 

 

page 11 of 27

 

Equipment & Furniture Component

 

   The Casing shall be constructed to withstand the standard operating Conditions
   in a Physician’s office and/or Hospital. The device must also with stand the
   harsh operating conditions associated with Ambulance use. The unit shall be safe to
   use both for the operator and the patient.
   Defibrillator
   The device shall utilize Biphasic waveform technology as the primary energy delivery
   option and shall include automatic and manual modes
   The biphasic waveform shall allow for escalating energies up to at least 200 Joules
   Safety: Internal discharge of energy if no discharge is effected.
   Internal self-testing and fault recognition.
   Required power 200-240V 50/60 Hz
   Weight not more 10kg
   Accessories :
   Full Standard Accessories Set-Up shall be supplied
   Standards
   CE;
   EC Marked
   US FDA;
   ISO certification
    
   Intensive care ventilator, for use on Pediatric/Infant- Adult patients
   With Graphics Display Monitor.
   Breath Types: VC, PC, VTPC
   Modes ( in each breath Type): A/CMV, SIMV, SPONT,
   Spontaneous Breath Choices: PS, VTPC.
   Features:
   Inspiratory time at least 0-3 sec with optional pause
   Inspiratory flow at least 3-140 l/min
   Pressure support: at least 0 to 45cm H2O/mbar
   Tidal Volume: at least 10-800ml
   Resp.Rate: at least 6-150b/min
   Pressure Limit: 0 to 80cmH2O/mbar
   I:E Rate: 1:4 to 4:1
   Trigger pressure or flow
   FiO2: 0.21 to 1.00
   PEEP/CPAP: 0 to 45 cmH2O/mbar
  2Leak Compensation
   Standby Condition: Allows setting to be preset and Circuit Check tests to be
   performed prior to starting Ventilation
   Open Exhalation Valve (BPRV): On/Off for Biphasic Pressure Release
   Ventilation
   Event History Log: Records 1000 events; alarm & settings
19Ventilator ICU Compliance Compensation: On/Off
 Graphical display of respiratory parameters
   
   Must be mounted over the Mobile Medical Air Compressor.
   Power requirements
   220 VAC ±10%,  50Hz.
   Internal rechargeable Battery provides an minimum. 60 minutes for complete
   ventilator function.
   Accessories :
   Medical Air Compressor
   Humidifier
   1 Tube set, adults
   1 Tube set, Children
   1 Humidifier chamber
   O2-connecting Tube 5.0 m
   1 Air-connecting Tube 5.0 m
   Humidifier : Heating power with safety thermostat and indication of operating status
   1 Circuit , Adults with patient Tubes , water traps, Y-Piece,mask elbow, catheter
   connector
   1 Circuit , Children with patient Tubes, water traps, Y-Piece,mask elbow, catheter
   connector
   1 Carriage
   Articulated Bracket for rail
   O2-cell without housing

 

 

page 12 of 27

 

Equipment & Furniture Component

 

   Sensors 5 pcs set
   Temperature Sensor
   Humidifier
   Bag- test-lung
   Standards
   CE;
   EC Marked
   US FDA;
   ISO certification
    
   Real-time SpO2%
   Oxygen Saturation Range (%SpO2)
   Measurement range at least 70-99%
   Accuracy (%SpO2) ±2% at100%~80%, ±3% at 80%~70%
   Real-time heart rate (BPM)
   Pulse
   Measurement range 30-250 bpm
   Accuracy ± 1 bpm at 30~250 bpm
   Display
   SpO2,Pulse rate, Pulse signal wave, Max/Min setup (SpO2% / BPM)
   Power Requirements internal rechargeable battery providing an operating time of at
   least 8 hours
   Compact design for easy transportability
20Pulsoximeter  
  2Accessories:
   3 rolls of paper
   Operation manual
   1 adult reusable finger Sensor
   1 Neonate Sensor
   Standards
   CE;
   EC Marked
   US FDA;
   ISO certification
    
   Patient monitor that shall draw Cardiograms, measures Blood Pressure (invasive or
  7not-invasive type), Respiration values, Temperatures at two spots, Oxygen Saturation
   in arterial blood ( SpO2 ), and Pulse rate ( HR ) .
   May be used to Monitor Patients on the transportation by using batteries, and the
   measured information on patient’s condition shall be displayed in wave forms along
   with numerical values.
   Display/Screen
   Display : Colour TFT display , screen diagonal Curves and measured values : not less
   than 25.00cm . Up to 6 Curves as well as numerical fields
   Brightness: can be manually adjusted
   Sweep Speed : Shall adjuste to 6.25, 12.5, 25 and 50 mm/s
   Display pixels: at least 1024×768
   Monitor Interfaces
   R-SYNC: for defibrillator , Each R-wave shall be clearly identified , Staff call:
21Patient Monitor, ICU Remote alarm via alarm adapter
 RS-232: as computer interface
   
   Mains/Battery Operation
   Mains operation: shall be via built-in Power Supply Battery operation: via integrated
   battery or Charging console . Battery operating time: approx. up to 4 hrs at least
   Power Supply : 220 V AC ±10%, automatic change-over Power failure buffer : for
   trend values,
   System Expansion
   Software updates via interfaces
   Monitoring parameters shall include : ECG, RESP, IBP, NIBP, SPO2, PULSE RATE
   , DUAL-TEMP
   S-T segment synchronized detection and analysis
   Graphic and tabular trends of all parameters for at least 96 hours
   20-Mn minimum ECG real time waveform and parameters storage and review
   Shall be able to be connected with a Central Monitor Station

 

 

page 13 of 27

 

 HSMP ArmeniaEquipment & Furniture ComponentHPIU: HH
    
   Inner high-definition thermal dot matrix recorder optional, which can output
   waveforms and characters 
   ECG : Input : Whole-lead ECG cable, standard AAMI cable connector
   Sweep speed: 12.5, 25, 50mm/s 0.5~40Hz(for monitor or operation)
   Calibration signal: 1mV , Protection: Against electrosurgical interference and
   defibrillation 
   Heart Rate: Measuring range: 30~250bpm , Alarm mode: Audible and
   visualAlarm 
   Respiraton Measurement  Range: at least 0~60 bpm 
   Oxygen Saturation(SPO2) Meas. Range: at least 70~100%
   NIBP : Method: Oscillometric , Mode: Manual, Auto , Resolution: 1 mmHg
   Shall have Over-pressure protection: (Adult 300mmHg ; Child 220mmHg
   Temperature : Meas. Range: at least 20~50C , Resolution: 0.1 C , Accuracy: +/- 0.2 C
   Operation Environment : Temperature: 0~45C , Humidity: 30~85%(non-condensing)
   Accessories 
   Electrode cable for disposable electrode (3 leads) 
   NIBP cuff for adults 
   NIBP cuff for children 
   Hose for NIBP cuff 
   Respiration Sensor 
   Power cord 
   Disposable electrode (5 pcs) 
   Temperature sensor – Adult (rectal) 
   SpO2 sensor reusable 
   Console / Bedside Holding Device 
   Standards 
   CE; 
   EC Marked 
   US FDA; 
   ISO certification 
    
   Patient monitor that shall draw Cardiograms, measures Blood Pressure (invasive or
   not-invasive type), Respiration values, Temperatures at two spots, Oxygen Saturation
   in arterial blood ( SpO2 ), and Pulse rate ( HR ) , End Tidal CO2 (etCO2) .
  2May be used to Monitor Patients on the transportation by using batteries, and the
  measured information on patient’s condition shall be displayed in wave forms along
   
   with numerical values. 
   Display/Screen 
   Display : Colour TFT display , screen diagonal Curves and measured values : not less
   than 25.00cm . Up to 6 Curves as well as numerical fields 
   Brightness: can be manually adjusted 
   Sweep Speed : Shall adjuste to 6.25, 12.5, 25 and 50 mm/s 
   Display pixels: at least 1024×768 
   Monitor Interfaces 
   R-SYNC: for defibrillator , Each R-wave shall be clearly identified , Staff call:
22Patient Monitor, anesthesia Remote alarm via alarm adapter 
 RS-232: as computer interface 
    
   Mains/Battery Operation 
   Mains operation: shall be via built-in Power Supply Battery operation: via integrated
   battery 
   or Charging console . Battery operating time: approx. up to 4 hrs at least
   Power Supply : 230 V AC ±10%, automatic change-over Power failure buffer : for
   trend values, 
   System Expansion 
   Software updates via interfaces 
   Monitoring parameters shall include : ECG, RESP, NIBP, SPO2, PULSE RATE ,
   DUAL-TEMP and etCO2 
   S-T segment synchronized detection and analysis 
   Graphic and tabular trends of all parameters for at least 96 hours
   20-Mn minimum ECG real time waveform and parameters storage and review

 

 

page 14 of 27

 

  Equipment & Furniture Component 
    
   Shall be able to be connected with a Central Monitor Station
   Inner high-definition thermal dot matrix recorder optional, which can output
   waveforms and characters 
   ECG : Input : Whole-lead ECG cable, standard AAMI cable connector
   Sweep speed: 12.5, 25, 50mm/s 0.5~40Hz(for monitor or operation)
   Calibration signal: 1mV , Protection: Against electrosurgical interference and
   defibrillation 
   Heart Rate: Measuring range: 30~250bpm , Alarm mode: Audible and visual Alarm
   Measurement Range: at least 0~60 bpm 
   Oxygen Saturation(SPO2) Meas. Range: at least 70~100%
   NIBP : Method: Oscillometric , Mode: Manual, Auto , Resolution: 1 mmHg
   Shall have Over-pressure protection: (Adult 300mmHg ; Child 220mmHg
   Temperature : Meas. Range: at least 20~50C , Resolution: 0.1 C , Accuracy: +/- 0.2 C
   End Tidal CO2 (etCO2) Module , Non Dispersive Infrared Spectroscopy Method
   Mainstream or sidestream 
   Displayed Data : Waveform labels and annotations ,etCO2, inC02 and respiration
   rate values 
   etCO2 Alarm Limits 
   Operation Environment : Temperature: 0~45C , Humidity: 30~85%(non-condensing)
   Accessories 
   Electrode cable for disposable electrode (3 leads) 
   NIBP cuff for adults 
   NIBP cuff for children 
   Hose for NIBP cuff 
   Respiration Sensor 
   Power cord 
   Disposable electrode (5 pcs) 
   Temperature sensor – Adult (rectal) 
   SpO2 sensor reusable 
   EtCO2 Mainstream or Sidestream accessories , Transducer , Lines , Filters
   Console / Bedside Holding Device 
   Standards 
   Classification: CF, defibrillation-proof, 
   CE; 
   EC Marked 
   US FDA; 
   ISO certification 
   Compact and easy to transport oxygen concentrator ( Mobile on Castors )
   The Oxygen Concentrator shall feature a dual-head Compressor. One head
   producing. Capacity : 1 to 5 l/Mn of O2 at 90% minimum at maximum flow
   Pressure-compensated Flowmeter shall permit use of long canula while maintaining
   accurate flow setting. 
   Safety Alarms: Audible and visual : Power Failure , Restricted Flow , , Low O2
   Pressure-relief Valve and thermal protection of the Compressor
   Double-insulated Unit , Two-prong plug 
   Flame-retardant Cabinet 
   Sound Level (ANSI): 50 dB average 
23Oxygen Concentrator Fixed humidifer Port and Recess shall prevent bottle and connector breakage
   
  1Power Requirements: 220 VAC,50 Hz 
    
   Accessories : 
   O2 Tubing , 3 Face Masks ( Adult , Infant , New Born ), Humidifier ,
   Set of Filters 
   Standards 
   CE; 
   EC Marked 
   US FDA; 

 

 

page 15 of 27

 

  Equipment & Furniture Component 
     
   ISO certification 
   Manufactured from chrome plated steel tube, three sections with independent
   manual controls and Trendelenburg position. 
24Couch, Gynecology4Top is upholstered with washable plastic material and flame retardant.
   Complete with padded Goepel leg-holders and S.S. bowl. 
   Knock down construction. 
   Manufactured from reinforced epoxy-painted steel tube 
   Three sections with independent manual controls. 
   Trendelenburg position. 
25Delivery Bed Third section extensible. 
  2Top is upholstered with washable plastic material flame retardant.
   Complete with S.S. tray, and two adjustable knee crutches, padded shoulder rest,
   handgrips, leg-holders. 
   Lamp , Examination . Mobile on 5 Castors , 5 star-shapedHeavy and stable Stand .
   Halogen Type 
   Height of LampHead shall be adjustable 
   Light Intensity shall be at least 15000 Lux with a Field radius of 17cm Minimum
   Color Intensity : 4300 °K. Single Reflector of 150mm Minimum
   Power rating : 1X40W Minimum . 220VAC/50Hz . Transformer , 24V DC shall be
26Examination Lamp3included 
 Accessories /Spares : 5 Halogen Bulbs 
    
   Standards 
   CE; 
   EC Marked 
   US FDA; 
   ISO certification 
27Uterine Aspiration Set1Stainless steel with handle for suction regulation. 
 Cannule ø mm. 6/8/10/12/14 
    
    
   CTG or Fetal monitor for recording and analyzing the fetal heart rate (FHR) on beat-
   to-beat basis and the uterine contraction with Doppler -US transducer and taco
   transducer. 
   Portable handheld CTG monitor. Compact and battery operated, shall allow Patients
   to be continuously monitored anywhere . Color display shall show the CTG data
  1presented in standard format for easy interpretation. 
   Fetal signals shall be easily located using the wide angle transducers with
   contractions stress being monitored by an external pressure transducer. Audio signals
   of each fetal heart are available on selection by the user through the built in Audio
   system. 
   The user will have control over the audio volume, audio fetal channel, chart recorder
   speed, toco zero, and event mark via the units front panel push buttons.
   Directional and watertight Ultrasound and TocoGraph with color-coded transducers
   for Single and Twins 
   Transducer: Multi element wide angle 
   Range: 30 to 240 bpm 
   Power output: < 20mW/cm2 
   Toco 
28Fetal Monitor Transducer: Real Time uterine Activity (contractions) trace on standard scale format
 Digital autocorrelation shall provide accurate and reliable FHR traces
   
   A pre-cut, non-fray elasticised belt with Buckle shall enable easy Transducer
   positioning for more accurate traces in difficult monitoring situations
   Clinical Event marker & automated foetal movement detection . Trace annotation
   Comprehensive patient database 
   Shall store over 500 patient records (30 minutes traces) in internal memory of FHR,
   Toco , Fœtal Movements with unlimited data storage on plug-in PCMCIA memory
   Cards 
   Computer interface 3 wire RS232 
   Thermal printer for common CTG paper 
   Speeds: 1,2,3, cm/min 
   Paper: z-fold pre-printed chart scale 
   The grid and the FHR/UA trace shall be printed simultaneously , ensuring printing
   accuracy and preventing potential errors in interpretation due to paper drifting or pen

 

 

page 16 of 27

 

  Equipment & Furniture Component 
     
   jitter 
   Power Supply :220 V AC and DC battery operated 
   Accessories: 
   1. Belts : 1 Original + 1 Spare Set 
   2. Buckles 
   3. IUP Starter Kit 
   4. Multicrystal transducer 
   5. Toco Transducers 
   6. Event marker Consumables: 
   7. CTG printer paper : 20 Packs 
   8. Fetal scalp electrodes 
   9. Ultrasound gel 
   10. IUP catheters 
   Standards 
   CE; 
   EC Marked 
   US FDA; 
   ISO certification 
   Capacity-0-15 kg 
29Scale for newborn children3Graduation-10 gr. 
Precision-+/- 10 gr. 
  
    
   Washable surface 
     
   Body with 6V 20W Halogen for 220V, 50 Hz. 
   Quadruple revolving nosepiece. 
   Sturdy,base with supportive rubber feet. 
   With low position, coaxial coarse and calibrated fine focus control, with graduation
   reading 2 microns per division. 
   Total focusing 40 mm. 
   Condenser adjustable. 
30Microscope Binocular1Built-in Power supply , halogen bulb adjustable. 
Double layer mechanical specimen stage. 
    
   Dust cover, blue filter, immersion oil and instruction manual included.
   2 pcs. 6V 20W halogen bulbs included. 
   Binocular Head, 30° inclined, 360° rotatable, Interpupillary Distance 55 – 75 mm
   Objectives: 
   ACHRO 4x/0.10 
   ACHRO 10x/0.2 
   ACHRO 40x/0.65, spring loaded 
   ACHRO 100x/1.25, Oil, spring loaded 
    
   Microscope with metal base with 360° rotatable monocular tube, 45° inclined
   3-fold objective nosepiece 
   separated coarse and fine focus 
   Stage focus control (protection of sample!) 
   Objectives: 
   ACHRO 4x/0.10 
   ACHRO 10x/0.25 
   ACHRO 40x/0.65 spring 
31Microscope Monocular1Objective 100x/1.25 oil, 
   attachable mechanical stage, etc. 
   Abbé condenser n.A. 1.25 with iris diaphragm, variable at height
   Stage with stage clips 
   Blue filter. 
   Built-in illumination 6V/20 Watt Halogen, adjustable 
   Daylight mirror for outdoor-use 
   Dust cover. 
   Packed in styrofoam case 
   Power Requirements: 220V / 50 Hz. 

 

 

page 17 of 27

 

Equipment & Furniture Component

 

To provide accurate and precise hematology results with WBC 3 parts differential. Principle method of detection: volumetric impedance, electrical impedance, laser flow cytometry, hydrodynamic focusing or equivalent.

 

It should be able to provide precise measurement of

  • White Blood Cells count (WBC): Precision (CV, coefficient of variation) < 3.5%
  • Red Blood Cells (RBC) : Precision (CV) < 2%
  • Platelets (PLT) : Precision (CV) <6.0%, linearity range: 10 – 950×109/L
  • Hemoglobin (Hgb)
  • Hematocrit (HCT)
  • Mean corpuscular volume (MCV)

 

  • Mean corpuscular hemoglobin (MCH)
  • Mean corpuscular hemoglobin concentration (MCHC)

 

  • Red blood cell distribution width coefficient of variation (RDW-CD) 1 10) Mean platelet volume (MPV)

 

  • Lymphocyte (L% & L#)
  • Monocyte (M% & M#)
  • Granulocyte (G% & G#)
  • Others

 

Please indicate all parameters included in the offer here Minimum WBC 3 parts differential

Minimum 3 histograms

 

32 Shall have dual sampling mode of both whole blood as well as capillary vascular Analyzer Hematology Efficient work flow control with patient selective testing

 

Shall be an open system to accept generic reagents and/ or local made reagents All reagents shall be cyanide free

With build-in printer

 

With quality control function and reporting

With programmable automatic cleaning and maintenance

 

Accept samples with any anticoagulant including EDTA, heparin and citrate acid Data storage of approximately 300 samples.

 

Approximately 60 samples/hour With operation menu, reporting

 

Possible with interfacing capabilities to external PC, Built-in Printer, LIS software or Network

 

The system offered shall be designed to operate normally under the conditions of the purchaser’s country. The conditions include Power Supply, Climate, Temperature, Humidity, etc.

 

Please declare in detail compliance of this item offered with any relevant quality and safety standards.

RS232

 

Power Supply : 220VAC.50Hz. Accessories :

Start-up Reagents as per Manufacturer

   Compact semi-automated urinalysis instrument with a throughput of 50 tests/h 
   Shall have individual setting of concentration ranges, grades and units 
   Automatic urine colour determination 
   Memory capacity of 150 results at least 
   Interface ports for connection to PC , bar-code reader and external printer 
   Results based upon test strips 
   Shall have long period calibration stability 
   Automatic compensation of intrinsic urine colour avoiding false positive results 
  1Power Supply : 
 33Urinalysis System220VAC.50Hz . 
 Accessories: 
    
   Strips , 1000 pcs 
   Calibration Testing Trips 
   Standards 
   CE; 
   EC Marked 
   US FDA; 
   ISO certification 
     
 34Chemistry AnalyzerAt least 100 tests/hr 
 Semi- AutomaticIncluding end point and kinetic measurement 
   
   Capable of analysing serum, plasma, urine and CSF 
     

 

 

page 18 of 27

 

Equipment & Furniture Component

Automatic calibration

Indication of abnormal values

Computer interface

Analysis of Albumin, ALP, ALT, AST, BUN, Ca++, Cholesterol, CK, Cl-, CO2,

Creatinine, Direct and total bilirubin, Glucose, Inorganic phosphorus, Iron, K+,

 

LDH, Mg, Na+, Total protein, Triglycerides, Urea, Uric acid and others

 

  1   
   Coagulometerwith 2 independent operating channels , Semi automatic operation 
   Tests performed: PT, APTT, Fibrinogen, cofactors II, V, VII, VIII, IX, X 
  1Optical detection with magnetic stirring used to detect clots 
   Temperature regulated at 37ºC ± 0.1ºC 
   Automated test calculation following addition of reagent, automatic compensation 
   for blanks  
   Straight forward input of test parameters, incubation period, tolerance for test 
   performed in double, reagent data (ISI, unit) 
   Automatic calibration/self test system 
   Capability to store calibration curves, presenting results in different units (second, 
   %, ratio, g/l or mg/ml. 
   Capacity PT: 60 in duplicate/per hour, APTT 45 in duplicate/per hour 
   LCD/LED display 
   Incubator capacity:  10 tubes 
   RS 232 interface for transferring results to PC or printer 
   Power supply:240VAC, 50Hz . 
   Accessories :  
   Thermoprinter  
35Coagulometer Starter kit consisting of 100 pcs of cuvettes with magnetic stirs 
   Consumables:  
   PT tests  
   APTT tests  
   Fibrinogen tests 
   Printer paper 20 meter rolls 
   Reagent cuvettes, 100 pcs 
   Magnetic stirrers, 10 pcs 
   Standards  
   CE;  
   EC Marked  
   US FDA;  
   ISO certification 
     
   Precise Temperature Control up to 99 ºC 
   Model with Capacities : 6 L 
   Seamless Stainless Steel Bath 
36Water Bath1Range Ambient +5ºC to 99ºC 
   Temperature Accuracy ± 0.5ºC 
   Uniformity ± 3.0ºC 
   Power Requirements: 220VAC / 50 Hz. 
     
   Productions rate: not less than 4 L/Hr 
   Standard equipment includes low water cut-off protecting heating elements from 
37Water Still2burnout when water supply is interrupted 
 Drainage Tap  
     
   Power Requirements: 220V / 50 Hz. 
     
   Centrifuges are required in the Laboratory to separate various components of Blood 
   for analysis.  
   Operational Requirements : 
38Centrifuge Table top version 
   Technical Specifications: 
   Tube Capacity :No. 24 – 36 : Size 5 – 15 ml 
   Should have a digital timer 
      

 

 

page 19 of 27

 

  Equipment & Furniture Component 
    
   Body should be made of strong fabricated & corrosion resistant steel
   Control panel – for start/stop switch, dynamic brakes, step less speed regulator with
  1zero start switch & speed indicator with timer and protective fuses.
   Door interlock 
   Maintenance-free brushless drive motor with exact speed pre-selection and display.
   Speed range 100 to 5000 rpm and above, accuracy 1 rpm.
   System Configuration Accessories, spares and consumables
   Centrifuge complete with Swing and basic rotors, 8 seats , and buckets
   Tube Holders as appropriate , 15 ml Tubes 
   Environmental factors : 
   Shall meet IEC-60601-1-2: 2001(Or Equivalent BIS) General Requirements of
   Safety for Electromagnetic Compatibility. 
   The unit shall be capable of being stored continuously in ambient temperature of 0-
   50 deg C and relative humidity of 15-90% 
   The unit shall be capable of operating continuously in ambient temperature of 10-40
   deg C and relative humidity of 80% 
   Power Supply : 
   Power input to be 220-240VAC, 50Hz . 
   Standards 
   CE; 
   EC Marked 
   US FDA; 
   ISO certification 
    
   Incubator 37° C – Outer case, inner case and front door manufactured with 18/10
   stainless steel, insulated with high grade glass wool.
   Door with silicon gasket and lock key. Front see-through inspection window.
   Heating elements shielded with 18/10 stainless steel.
   Adjustable mesh shelves manufactured with 18/10 stainless steel. Forced air
39Incubator laboratory2ventilation. 
Thermostat 
    
   Thermometer with analogical dial. Safety thermostat.
   Possibility of continuous running : 0-120 min. 
   Capacity not less than 60 liters 
   Power Requirements – 220 ± 10% V. 50 Hz. 
     
   For medicine storage. 
   Upright Type 
   Internal volume not less than 250 lt. / freezer – 40÷50 lt.
40Refrigerator3Temperature controlled within the range of +2°÷ +8° C.
  Number of shelves – 3 or 4. 
    
   With defrosting system 
   Power Requirements: 220 ± 10% V. 50 Hz. 
   Horizontal autoclave, benchtop Type for CSSD,Laboratory , OutPatient/Wards
   Sections 
   Self-contained ,front-loading and fully automatic Unit with integrated Steam
   Generator . 
   The Unit shall be single Manual Door Type 
   Stainless Steel Chamber and Door : Z3CN18/10 
   Capacity per cycle shall be at least 100L for a Cycle time of not more than 25 Mn.(
   Sterilization Phase at 134°C). 
   Sterilization shall be enhanced by the use of Vacuum Cycle Digital temperature
   control and timing , Pressure Control , fully programmable cycle parameters
41Autoclave Low Water alarm, Door Safety Locks, Program Cut-Out in case of Power Failure ,
   Automatic Zeroing. Cycle End Alarm 
  2During Cycle progress , Door is locked into position by both Temperature and
   Pressure Interlocks . Rapid single action door closure shall be of concern
   Electronically-operated Vent valve , Automatic Air Purging , Over Pressure Safety
   Valve and Low Water Cut-Out with Visual and Audible Alarm indication shall be as
   standard 
   The Unit shall be equiped with at least 4 Stainless Steel Trays
   Power Supply: 220VAC.50Hz . 
   Accessories : 
   4 Stainless Steel Trays 

 

 

 

page 20 of 27

 

  Equipment & Furniture Component 
     
   Standards 
   CE; 
   EC Marked 
   US FDA; 
   ISO certification 
   Electronic Controlled Sterilizing Process 
   Outer case and internal chamber – constructed from stainless steel.
   Two or three adjustable shelves. 
   Wall and hinged Door Insulated with High Grade Glass Wool.
42Hot air sterilizer3Door with Lock and Flat Key. 
 Temperature: adjustable from +50ºC to +200 ºC ±10 %. 
    
   Timer Range: 0-250 minutes. ±10 %. 
   Capacity: 50- 60l. 
   Power requirements 
   220 VAC ±10 %,  50Hz. 
 Scale for Adults1Capacity-5-150kg. 
43Graduation-100gr. 
   
   Precision-25 gr. 
   Washable surface. 
   Manual Resuscitator, Autoclavable,with unidirectional Valve
   The supply of Filters shall permit the use in Toxic Atmosphere
   Adapted to First Emergencies the unit shall also the complement to all automatic
   Ventilation 
   Designed for Adults . Shall be supplied with 2 Masks 
   All parts shall be autoclavable up to 134° C. 
   The device shall consist of a self-filling Bag of thin Silicone Rubber which shall
   transmit adequately the User’s Fingertips Action with a high degree of ‘feel’
   indicative of Pressure and Volumes Changes 
   An inflating valve ( Transparent Polysulfone ) shall be fitted between the orifice of
   the Bag’s Delivery End and 2 Rubber Face Masks shall be supplied
   Designed Leak shall counteract Inflating Pressure for all Lung’s Conditions
   A pressure relief valve shall be incorporated to comply with BS 6850.1987
   Balloon : Silicon 1,5L 
  1Safety Pressure : 30 cm H2O 
44AMBU emergency caseAccessories : 
  
   Balloon , autoclavable 
   Valve, autoclavable 
   Valve Obturator 
   PEP Valve Adaptor 
   2 Masks vinyl, Adult / Children with Straps 
   Tube for Pressure retake 
   Plastic carrying case 
   Standards: 
   CE; 
   EC Marked 
   US FDA; 
   ISO certification 
   Bacterial Lamp for Air Sterilization by means of UV Beams , UVC Category or Far
   UV (200nm< >280nm) with Strong bactericide effectiveness with a maximum
   strength situated at 265nm (253.7nm Mercury Spectra ) 
   UVC shall be emitted thru vacuum ionizing Tube ( Weak permeation )
   Useful power shall be at least 45% of nominal power (30W)
   LifeTime shall be at least 7500 Hrs 
45Bacteriological light15Radiation shall be witnessed by means of blue Light emission
Encased in Steel Sheet casing , Epoxy-painted. Ceiling or wall fixation
   Power supply:220V.AC50 Hz. 
   Accessories/Spare Parts: 
   Ionized Tube : 5 
     
46Infusion fluid holder   

 

 

page 21 of 27

 

  Equipment & Furniture Component 
      
  20IV Pole , Mobile on Castors  
   Adjustable height  
   2 Hooks  
     
   Manufactured from chrome plated steel tube 
   Upholstered fixed Top covered with washable plastic material , Flame Retardant
   Complete with side Rails and IV Stand 
47Patient trolley5Mounted on 4 casters 100 mm two of which with brakes 
  Knock down construction.  
     
   Length min. –2000 mm.  
   Height – 700 mm.  
   Nursing incubator mobile on castors 
   Temperature control by air temperature or skin temperature 
   Temperature control range 20-37 °C 
   Display of temperature  
   Alarms  
   High and low air temperature  
   High and low skin temperature  
48Incubator neonate Fan failure  
 Sensor failure  
   Power failure  
   Secondary thermostat  
   Heater power indicator  
   At least 2 hand ports  
   At least 2 tubing ports  
   Provision for oxygen supplementation 
   Provision for humidity supplementation 
   Internal noise < 60 dB(A)  
   Power supply 220 V 50 Hz  
   Mobile, Stand-type.Max. Light intensity (Ec) up to not less than 110
   000 lx or more Color rendering index (general) Ra 94-96 Color
   rendering index (red) R9> 90Light field size approx. 13-25 cm. Co lour
49.Surgical light mobile temperature 4200º K (approx.) Working range without focusing
 intensity approx.70 – 130 cmPower supply 220 V, 50 Hz Light source 3
   
   x halogen approx. 23 V /50 WDurability of halogen bulbs 1300 h or
   more Adjustment of light intensity approx.50 – 100% (0-100%)
   Sterilizable handle  
50.Splint set Available  
51.Transportation splint Available  
52.Stretcher Available  
54.Tracheostomy set Available  
   Optical System  
   Field of view: 100°÷ 120°  
   Depth of field: 3÷50 mm.  
   Distal end  
   Bending Section  
   Range of tip bending: not less than Up 180°, Down 130°. 
   Insertion tube  
   Outer diameter: not more than 5.0mm 
   Working length: not less than 600mm. 
   Total lenhgth: not more than 900mm. 
55.Fiber bronchoscope Instrument channel  
 Inner diameter: not less than 2.0mm. 
    
   Light Source  
   Mobile  
   Lamp  
   Type:Quartz halogen.  
   Output: 150W.  
   Cooling fan prolongs lamp life not less than 500 hours. 
   Air/Water functions available  
   Power consumption:not more than 250 W. 
   Power Requirements – 220 ± 10% V. 50 Hz. standard network
   Weight: not more than 6 kg.  
      

 

 

page 22 of 27

 

  Equipment & Furniture Component 
    
   Telescope: 1.2mm , 0º direction of view. Trocar Tube: 1.2mm Trocar
   spike: 1.2mm Accessories: Adapter with Luer lock male and x-mas tree
   Telescope holder Light quid, 3.5m with condenser Trocar Tube for
56.Arthroscope surgical instruments 2mm Trocar spike 2mm pointed Trocar spike 2mm
 blunt Palpation Probe 2 mm straight -4pcs. Palpation Probe 2mm Hook
   
   Type -4pcs. Knife 2mm side cutting -4pcs. Knife 2mm front cutting –
   4pcs. Rasp 2mm -4pcs. Biopsy Forceps 2mm -4pcs. Grasping Forceps
   2mm -4pcs. Instruments Tray Shaver System
   Optical System Field of View: 80º minimum Direction of View: Lateral,
   Rear-view 15º minimum Depth of Field: 5-60 mm minimum Distal End
   Outer Diameter: max. 12.0mm Instrument Channel Inner Diameter:
57.Duodenoscope minimum  2.8mm  Insertion  Tube  Outer  Diameter:  max.  11.0mm
   Working Length: minimum 1235mm Bending Section Range of Tip
   Bending minimum: UP 120º, Down 90º Right 110º, Left 90º Biopsy
   Forceps Minimum Visible Distance: 10 mm from Distal End
   Laparoscopy Surgery Set Including:
   Forward-Oblique Telescope 300 deg, diameter 10mm, length 31cm,
   including 
   fiber optic light transmission incorporated. Color code: red.
   Forward-Oblique Telescope 300 deg, diameter 5mm, length 29 cm,
   including 
   fiber optic light transmission incorporated. Color code: red.
   VERESS Pneumoperitoneum Needle with spring loaded stylet LUER-lock,
   length 13 cm. 
   Trocar 11mm with Cannula and multifunctional valve.
   Trocar 6mm with Cannula and multifunctional valve.
   Reducer 11/5mm. 
   Reduction Sleeve, reusable, instrument diameter 3mm, trocar cannula O.D.
   6 mm. 
   Reduction Sleeve, reusable, instrument diameter 10mm, trocar cannula O.D.
   15 mm. 
   Reduction Sleeve, reusable, instrument diameter 5mm,trocar cannula O.D.
   11 mm.color code:green 
   Dissecting and Grasping Forceps, rotating, with connector for unipolar
   coagulation size 5mm, length 36 cm.
   KELLY and Grasping Forceps, rotating, with connector for unipolar
   coagulation size 5mm, length 36 cm.
 Laparoscopy system with KELLY Forceps insert, long 
58.instrumentation for MANHES grasping forceps with connector for unipolar
 general surgery coagulation 5mm, length 36cm 
   METZENBAUM scissors, rotating, with connector for unipolar
   coagulation 5mm, length 36cm,blades curved, double action jaws, length of
   blades 12 mm. 
   Insert METZENBAUM scissors 
   Hook Scissors, rotating, size 5mm, length 36cm,single action jaws
   Coagulating and Dissecting Electrode, L-shaped, with connector pin for
   unipolar coagulation size 5mm, length 36 cm
   Coagulating and Dissecting Electrode, spatula-sheaped, blunt with
   connector pin for unipolar coagulation size 5mm working length
   Insert Micro-dissecting scissors 
   BLAKSSLEY Forceps insert 
   Suction and coagulation cannula, L-shaper
   Take-apart Bipolar Coagulating Forceps
   Insert Forceps 
   Applier for ligating clips (medium large) dismanting, rotating
   Pilling-Titanium-Clips, medium-large, sterile box with 16 cartridges,
   10clipseach 
   Electronic CO2 Endoflator, CO2 bottle empty, power supply, 240 VAC, 50
   Hz, 
   CO2/ N2O gas filter for use withElectronic Endoflator
   High Frequency Surgery Unit, power supply: 200-240 VAC, including: 400
   A Mains Cord Adaptor, unipolar HF- output, 4 mm/5 mm

 

 

page 23 of 27

 

  Equipment & Furniture Component 
    
   Connecting Cord, for connecting the neutral electrodes
   Neutral-Electrode made of conductive silicone, with 1 rubber strap for
   fastening contact face, surface area approx. 180 sq.cm.
   Two-Pedal Footswitch, for control of coagulation.
   Monopolar High Frequency Cord with 5 mm plug for HF-unit, length 300
   cm 
   Suction Bottle, 1.5L. 
   Bacterial filter 
   Cold Light Fountain: approx. 175 W power supply: 220-240 VAC, 50 Hz
   Fiber Optic Light Cable, size: 4.8 mm, length 180 cm
   Endoscopic video system, color system PAL power supply: 240 VAC, 50
   Hz. 
   Cover for use with endoscopic video cameras, size 17 x 242 cm, rolled up.
   Sterile, in packs of 15. 
   Color Monitor, color system PAL, NTSC screen 45 cm power supply: 240
   VAC, 50 Hz including: mains cord 
   Mobile  cart  for  video  system:  4  antistatic  dual  wheels,  2
   equipped with locking brakes, 2 fixes shelves, 1 drawer unit with lock, 1 set
   of non-sliding stands for units, inclusive integrated small cable conduit in
   both vertical boom, 1 camera mount.  Power Box socket board with 12
   plugs,  12  grounding  plugs  Approximate  Dimensions:  cart:  700mm  x
   1280mm x 686mm (wxhxd) Shelf 630mm x 480mm (wxd) caster diameter:
   125 mm 
    
   Electroencephalograph- analyzer. Based on the personal computer. Routine
   EEG studies with computerized analysis and topographical mapping, in-
   depth analysis of EEG/EP activity, analysis ofepileptiform activity, long-
   term EEG monitoring , differential diagnosis24-EEG inputs. Sensitivity:
   0.1 µV to 200 µV. Calibration µV: square (1Hz.) or Sine (5Hz) signal of 5,
59.EEG 10 ,20, 50, 100, 200 and more High frequency filter: >70 Hz at highest
 setting) Low frequency filter: <0,3 Hz at lowest setting) AC Rejection
   
   filters (50/60 Hz.): 1:1000 or better. Automatic suppression of miographical
   artifacts: 20 dB (30 Hz.) or better. Digital signal processing: 22-bit ADC
   Software-controlled Photo and Audio stimulators. Check-up of electrode
   impendence. Power Requirements – 220 ± 10% V. 50 Hz. standard network
   Standard set . Set of standard accessories. EEG-gel 2 kg.
   Gantry 
   Number of slices per rotation-2 
   Gantry aperture: at least 65 cm 
   Gantry inclination angle: at least. 20 degrees 
   Complete scanning time: 1.5, 2.0, 3.0, 5.0 sec app. 
   Min. scanning time: app. 1.5 sec 
   Patient table 
   Coverage region: 70 cm 
   Table height: 40-90 cm 
   Scanning horizontal range: 120 cm 
   Max patient weight: app. 180 kg 
   Acuracy of horizontal movement +/-: 0,25 mm 
   Detector 
60.CT-scanner Detector type: Solid body 
   Number of detecting elements: at least 708 per row 
   High-contrast resolution: at least 13 par. line/cm at 0% MTF
   Low-contrast resolution: less than 4 mm at 0,3% at 20 mGy
   X-ray subsystem 
   X-ray generator capacity: min 24kWt 
   mA range at least 20-250 mA 
   kVp range at least 80-140 kV 
   Scanning parameters 
   Max duration of spiral scanning: 60 sec 
   Pitch: 0,1:1 to 3,0:1 
   Slice thickness: 1,2,3,5,7,10 mm 
   Min slice thickness: 1mm 
   Summary volume of hard discs: min 80GB 
   Monitor: LCD 

 

 

page 24 of 27

 

  Equipment & Furniture Component 
     
   Monitor size: min 17’’ 
   Power requirements: 380V, 3-phase. 
    
   General Description: General purpose portable ultrasound scanner, with
   multiple focus electronic probes. 1. System Specifications: PROBES:
   Electronic 3.5MHz convex Electronic 7.5 MHz linear Scan Modes: B, M,
   B+M, B+B Scan Converter: −Matrix size: min 512x512x8 −Gray-scale
   levels: min 256 Software Features: −Multiple beam focusing −Image
 Portable ultrasound preprocessing −Image postprocessing −Zoom −Digital calipers −General
61. analysis: distance, area, circumference −Obstetrics software
scanner 
  packageKeyboard: Alphanumeric keyboard withpointing device (joystick
   
   or trackball). Maximum Display Depth, cm: 22 or more Monitor: High
   resolution monochromatic, minimum size: 7-inch Connections: -standard
   video output 625 lines / 50 Hz -image freezing footswitch Power
   Requirements: -Mains: 220V, 50Hz -Rechargeable battery, minimum
   capacity: 1 hour WEIGHT: maximum 15 kg 
   Type : Galilean Magnification : 4.5X, 7.5X, 11.5X, 18.5X, 29.5X or similar
   Field of View : Ø 49 mm to Ø 8 mm Illumination : Achromatic shadow
62.Surgical microscope for free spot Light Source : Cold light through fiber-optic light guide Bulb :
eye Reflector type halogen bulb 15V 150W (24V 250W optional) Illumination
  
   field : 48 mm (Ø) Light intensity : 80000 LUX max. Filter : Various filters
   are available 
   Infant Warmer for Intensive Care Procedures, Includes Micro-processed
   Temperature Control and Orientable Halogen Examination Lamp, with
   Lateral Movement Allowing Free Access for X-Ray.
   Crib of Wide Dimensions 55×65 cm approx. with Washable and Hypo-
   allergic Mattress Tray Apt for X-Rays. 
   Lateral Panels Foldable up to 180º, with Four back IV Ports.
   Trendelenburg and Fowler Movement of ±15º minimum
   Displays: 
   Digital Skin Temperature Thermometer 
   Resolution ≥ 0.1ºC, Range minimum 20 to 42 ºC 
   Temperature Accuracy ±0.3ºC or better 
   Digital Set Point, Resolution 0.1 ºC, Controllable Range minimum 34 ºC to
   38 ºC 
   Heater Power Indicator 
   Skin or Manual Control Mode Indicator 
   Performance 
   -Continuous Circuit Monitoring 
   -Self-calibrating Temperature Circuits 
   -Automatic default Set-point of 36 ºC for Skin Temperature and
   Memory for Control Temperature 
63.Infant reanimation centre -Memory for Control Set-point Short Term Energy Failure
   -Interchangeable Skin Temperature Probes: ±0.1ºC
   Alarms 
   Skin Temperature ±1ºC 
   Patient Alarm: Heater Remains at max. Power for more than 15
   Consecutive minutes. 
   Skin Temperature Sensor Failure. 
   Electronic Circuit Failure 
   Power Failure 
   Alarm Reset and other 
   Accessories: 
   Skin Temperature Sensor 
   Lateral Rail for Accessories 
   IV Pole 
   X-Ray Cassette Tray 
   Patient Identification Card Holder 
   Monitor Shelf 
   Auxiliary Connection Box 
   Holder for Puls-Oximeter 
   Puls-Oximeter for Infants 
   Lateral Handles for Moving and Instruction Manual

 

 

page 25 of 27

 

  Equipment & Furniture Component 
     
   Phototherapy Lamp 
   Resuscitation System 
   Power requirements 
   220 VAC ±10 % , ≤50W, 50Hz. 
   Twins monitoring capability 
   External ultrasound transducers for measurement of foetal heart rate
   Detection of foetal heart echo by autocorrelation 
   Foetal heart rate tange at least 50–210 /min 
64Cardiotocograph Alarms for high and low foetal heart rate 
   External displacement transducer for measurement of uterine activity
   Digital and graphical display of foetal heart rates and unterine activity
   Two channel recorder with event marker and annotation capability
   Power requirement 220 V, 50 Hz 
   Nebulizer with compressor on mobile stand 
   Aerosol particles in range 1 to 5 micron 
65Nebulizer Flow rate at least 4 ml/minute 
   Liquid/medication container at least 200ml 
   Power requirement 220 V, 50 Hz 
   High frequency generator 
   kV range at least 22–35 kV with 1 kV increments 
   mA range up to at least 100 mA 
   Time range at least 0.05–86sec 
   mAs range at least 5–500 mAs 
   Automatic exposure control 
   X-ray Tube 
66.Mammograph rotating anode 
   heat capacity at least 300,000 HU 
   heat dissipation rate at least 60,000 HU/min 
   target/filter Mo/Mo, Mo/Rh 
   dual focus 0.1/0.3 mm 
   Breast compression system with automatic, manual and fine adjustment
   Grid ratio 5:1 
   Radiation shield with lead glass as required 
   Single plane angiography system 
   4-way table top motion 
   Collision prevention system 
   X-ray Generator 
   high frequency 
   100 kW @ 100 kVp 
   radiographic mA up to 1000 mA 
   radiographic kVp 60–150 kVp 
   radiographic timer up to 1.5 sec 
   fluoroscopic mA up to 10 mA 
   fluoroscopic kVp 60–110 kVp 
   pulsed fluoroscopy 
   cine range up to at least 30 f/sec 
 Coronarography/Angiogr X-ray tube 
67. 80 kW 
aphy system  
  heat storage capacity at least 1,500 kHU 
    
   heat dissipation rate at least 300 kHU/min 
   dual focus, focal spot size 0.6/1.2 mm or less 
   grid pulsed and cooled 
   Digital dynamic flat detector 
   23 cm with zoom fields 
   Storage capacity at least 50,000 images at 1,024 × 1,024
   At least two 17 inch monitors 
   Digital subtraction angiography 
   Cardiac analysis software package 
   Other software packages to suit user applications 
   DICOM 3 compatibility 
   Radiation shield as required 
   Contrast injector 

 

 

page 26 of 27

 

  Equipment & Furniture Component 
     
   Monitoring system 
   Accessories as specified by user 
   Room preparation as required by user 
   Convertible into operating table for emergency Caesarian section
   Adjustable backrest 
   Removable headboard 
   Leg supports 
68.Delivery Couch Removable foot section 
   Height adjustable over at least 60-90 cm 
   With castors at least 12 cm dia, at least two diagonally lockable
   Corner bumpers 
   Removable drain pan 
   Heat control 
   manual or automatic 
   range includes 34–37 °C 
   heater indicator 
   Temperature display 
   set point temperature 
69Infant warmer skin temperature 
   range includes 30–39 °C 
   Alarms 
   audiovisual 
   sensor disconnect 
   power failure 
   Power requirement 220 V, 50 Hz 
   Capacity at least 50 kg/cycle 
   Cylinder volume at least 500 litre 
   Stainless steel construction of parts exposed to corrosion
70.Washing machine Electric heating 
   Adjustable extraction speed including 300–500 rpm 
   Multiple liquid supply and detergent connections 
   Programmable 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

page 27 of 27